40
SM6016
9.
Performance Check
Accuracy Specifications
Notice
1
.
Environment temperature
:
23
℃
± 5
℃
; Humidity:
≤
75% R.H.
2
.
Valid after 10 minutes of warm up time.
3
.
Test in measuring slots on front panel.
4
.
Measurements performed after correct open and short correction.
5
.
Test in the recommended equivalent mode.
6
.
Percentage discrepancy Indication: ±
(
% number of least significant
digits
)
7
.
Actual measurement and display range exceed the ranges specified in table, but
we do not assign accuracy for measurement values exceeding these ranges.
8
.
When the test level is 0.3V: accuracy*2
9
.
Subscript Explanation: S—series equivalent; p—parallel equivalent; e: accuracy
10
.
Some parameters cannot be expressed as the way of data sheet; therefore
they can just be converted by formulas according to corresponding results.
Inductance (L) and Quality Factor (Q)
Accuracy
Range
Display Range
Le
De*
Recommended
Equivalent
Mode
1000H
400.0H - 1000.0H
1.00%+3 digits
0.0100
Parallel
400H
40.00H - 399.99H
0.35%+2 digits
0.0035
Parallel
40H
4.000H - 39.999H
0.25%+2 digits
0.0025
Parallel
4H
400.0mH - 3.9999H
0.25%+2 digits
0.0025
----
400mH
40.00mH - 399.99mH 0.25%+2 digits
0.0025
Series
40mH
4.000mH - 39.999
0.45%+2 digits
0.0045
Series
1
0
0
H
z/
1
2
0
H
z
4mH
0uH - 3.999mH
1.40%+5 digits
------
Series
100H
40.00H - 100.00H
1.00%+3 digits
0.0100
Parallel
40H
4.000H - 39.999H
0.35%+2 digits
0.0035
Parallel
4H
400.0mH - 3.9999H
0.25%+2 digits
0.0025
Parallel
400mH
40.00mH - 399.99
0.25%+2 digits
0.0025
----
40mH
4.000mH - 39.999mH 0.25%+2 digits
0.0025
Series
4mH
400.0uH - 3.9999
0.45%+2 digits
0.0045
Series
1
k
H
z
400
µ
H
0.0uH - 399.9
µ
H
1.40%+5 digits
------
Series
1000mH
400.0mH 999.99 mH
0.80%+3 digits
0.0080
Parallel
400mH
40.00mH 399.99 mH
0.35%+2 digits
0.0035
Parallel
40mH
4.000mH 39.999 mH
0.25%+2 digits
0.0025
----
1
0
k
H
z
4mH
400.0uH 3.9999 mH
0.30%+2 digits
0.0030
Series