SM6016
29
7.3 Capacitance Measurement
WARNING: Before testing, ensure that the tested capacitor has been
fully discharged.
1. Press the POWER button for a long time to turn on the meter.
2. Press the PRI button until “C” is displayed on the screen to select
capacitance measurement.
3. Insert a capacitor into test slots or connect a tested capacitor through a
proper test accessory (e.g., banana plug-crocodile clip test leads, test
fixture or SMD test tweezers). See figures 14 below. Figure 5 shows the
optional 4-terminal test accessories.
4. CAUTION: BEFORE inserting a capacitor or capacitive component
into the input slots or terminals, be sure to fully discharge the
component. Some larger capacitive components may take longer to
discharge. In these cases, please allot enough time for a full
discharge. If proper discharging of the
component is not done correctly, it will damage the input terminals
of the meter.
5. Press the FREQ button until the desired test frequency is displayed on
screen.
6. Press the LEV button until the desired test level is displayed on screen.
7. Press the SEC button to select the desired secondary parameter.
8. Read the readings on LCD for capacitance measured values.
Figure 14- Capacitance Measurement