SM6016
Command Reference
Common Command
*IDN?
Query instrument ID.
Return: <instrument model>, <firmware version>, <serial number>
*LLO
Local Lockout. This means that all front panel buttons, including the RMT key is not
available. (POWER button is enabled.)
*GTL
Go to local and remove local lockout. If *LLO is sent, the only way to operate front
panel is to go to *GTL.
*TRG
Trigger the instrument to take a measurement. Due to the automatically continuous
test, *TRG command is of no use.
SCPI Commands
FREQuency Subsystem commands
FREQuency <value>
Description: Set the test frequency Parameters: 100, 120, 1000, 10000,100000 or
100Hz,120Hz,1kHz,10kHz,100kHz
Example: FREQuency 100Hz
Set the frequency to 100Hz
FREQuency?
Description: Query the current test frequency
Return: <100Hz|120Hz|1kHz|10kHz|100kHz>
VOLTage subsystem
VOLTage <value>
Description
:
set the test level (only effective in L,C,R,Z)
Parameters are 0.3, 0.6, 1 or 3e-1, 6e-1,1e0
Example
:
VOLTage 0.3
Set the test level to 0.3 V
VOLTage?
Description: set the current test level
Return
:
<0.3V|0.6V|1V>
FUNCtion subsystem
FUNCtion:impa < L | C | R | Z | DCR >
Description: Select the primary parameter
Example: FUNCtion:impa L
Selects L as the primary parameter
FUNCtion:impa?
Description: Query the primary parameter
Return: <L, C, R, Z, DCR, NULL >