CALCulate:RECording:MAXimum?
Description: Query the maximum value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds limits
or there is no data, what returns is“----”.)
CALCulate:RECording:MINimum?
Description: Query the minimum value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds limits
or there is no data, what returns is“----”.)
CALCulate:RECording:AVERage?
Description: Query the average value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds
limits or there is no data, what returns is“----”.)
CALCulate:RECording:PRESent?
Description: Query the present value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds limits
or there is no data, what returns is“----”.)
FETCh Subsystem
FETCh?
Description: Returns the primary, secondary display value and tolerance compared
result (BIN no.).
Return: <NR3, NR3, NR1> when the primary parameter is LCR, Primary parameter,
secondary parameter and BIN no.
<NR3,NR1> when the primary parameter is DCR, Primary parameter and BIN
no.
Example: FETCh?
Summary of Supported SCPI Commands
Table 7 - Summary of SCPI Commands
Command
Parameter
Explanation
FREQuency
<Value>
Set the Test Frequency
FREQuency?
Query the Test Frequency
VOLTage
<Value>
Set the Test level
VOLTage?
Query the Test level
FUNCtion
:impa
<Literal>
Select the primary display parameter
:impa?
Query the primary display parameter
:impb
<Literal>
Select the secondary display parameter
:impb?
Query the secondary display parameter