3
1.1.3 Lamp Replacement
1.
Turn the power switch to the “Off” position.
2.
Remove the fiber optic component.
3.
Allow the lamp to cool.
4.
Unscrew the “Modulamp” assembly captive thumbscrew until the head springs
forward (H).
5.
Pull the top of the “Modulamp” assembly forward and lift the unit out of the
recess (J). Do not pull the assembly beyond the length of the cable assembly
(M).
CAUTION: Lamp, lamp socket and surrounding surfaces may be hot!
6.
Push back on the lampholder release lever to raise the lamp (L).
7.
Gently unplug the ceramic lamp socket (K).
8.
Lift the release lever. Insert a new lamp into the lamp holder. Push the lamp
all the way down. Be careful not to touch the bulb or the inside of the
reflector. Fingerprints may affect the light output.
9.
Push the ceramic socket on to the two lamp pins until flush with the lamp
base.
10.
Return the “Modulamp” assembly into the recess and tighten captive
thumbscrew.
11.
For complete “Modulamp” assembly removal and installation, see Sections 5.1
and 5.2.