12. Trouble Shooting
If the EndoPilot does not seem to work properly, it does not necessarily have to be a device error. First, please check
device with the help of the following table, in order to exclude any handling errors or disturbances (like, for example,
anatomic peculiarities).
Problem Possible
Reason
Solution
Device in general
Device shows no function
and display does not work.
No power
Is the power supply connected correctly (LED must be
shining)?
Plug in the D-Pack hand-piece only after turning the device
on.
Use of Touch-Display not
possible, device does not
react.
Touch of display damaged Result of using excessive force.
Contact the manufacturer.
No acoustic signals
Tones are turned off
Turn the tones on again.
Foot switch without
function
Foot switch damaged
Check the cables and the plug for damages.
EndoMotor
Instrument does not turn.
Calibration not been done Execute calibration with contra-angle.
Motor damaged
Check the cables and the plug for damages.
Check if the motor is running with contra-angle.
Contra-angle damaged
Check if the axle can be turned freely.
Apexlocator
No measurement possible Contact problems
Are lip clip and cable connected correctly?
Are lip clip and file clamp absolutely clean?
Make sure that a short circuit is shown on the display when
file contact and lip clip are touching each other.
Lip clip at the wrong
connection
The lip clip must be connected to the female connector of
the Apex-cable (socket)!
Wrong contra-angle
Please check if the EndoPilot contra-angle has been
connected. Did it snap in correctly? Bring lip clip and NiTi-
file in touch with each other. Does the display show a short
circuit?
Root canal calcified or
obliterated
Check x-ray photo, eventually catheterisation with suitable
file until working length.
Root canal very dry
Rinse with NaCl. Dry access cavity with cotton pellet.
Blocking by old filling /
medicated substances
X-ray for comparison! Complete removal of old
Guttapercha remnants or of medicated substances.
DownPack
Instrument does not heat
Instrument damaged
Insert new heating tip.
Instrument is turning
Use the tool to tighten the nut.
BackFill
No Guttapercha at the
needle
Hand-piece too cold
Was the heating time long enough? Is the heating chamber
warm?
Pellet is used up
Is the piston completely in front position?
Insert a new pellet.
Instrument turns
Turn the knob clockwise until the stop (see 2.9)
Piston is blocked (cannot
be pulled back during
cleaning)
Guttapercha remnants
The remnants stick to the piston, the gun has to be emptied
completely while it is still warm.
Ultrasonic
Low performance
Tip not tightened,
deformed or worn out
Tighten instrument with tool
Exchange instrument
No function
Tip not tightened, cable
eventually damaged
Tighten instrument with tool
Contact your supplier or Schlumbohm GmbH & Co. KG
Wireless foot switch
No function
Batteries empty
Open the battery compartment at the ground plate of the
foot switch and change batteries
No function
Switch not recognized
The wireless-switch should be paired / see chapter 8
If the problem cannot be solved, please contact your supplier or Schlumbohm GmbH & Co. KG directly.