4.7. Options
The settings, which will be valid for the entire treatment
can be selected in the menu O p t i o n s .
The setting of the curvature of the root canal serves
for the calculation of the wear.
The following can be selected as r o o t - c u r v a t u r e :
easily
(Factor 1x)
medium
(Factor 2x) – pre-set
strong
(Factor 3x).
A p e x - f u n c t i o n s , here you can select (see chapter 4.7.1):
auto-stop
(arbitrary, stopping time: 0,5, 1, or 2 seconds)
no auto-stop
off
a u d i b l e a s s i s t a n c e , here you can select:
off
closer to apex
- the closer the file gets to the Apex,
the shorter the time between the tones.
Motor Load
- the stronger the torque-load (file is approaching the
maximum torque limit), the shorter the times between the tones.
The function 1 : 1 c a l i b r a t e c o n t r a - a n g l e so
should be used after each change and each sterilization of the
contra-angle.
This can compensate smaller torque-losses at the contra-angle and helps to ensure a stable usage of
small torque-limits. If the calibration is not possible, the contra-angle is dirty or damaged. In this case, please contact the
manufacturer.