Operating manual for the VISATRON VN2020
V2.0 12/2021 EN
Page
69
of
106
SCHALLER AUTOMATION
66440 Blieskastel / Saarland / Germany / Industriering 14 / P.O. Box 1280
Industrielle Automationstechnik GmbH & Co. KG
T
el. +49(0)6842-508
-
0
/ Fax -260 / Email: [email protected] /
www.schaller-automation.com
3.
Figure 104: Unplug plug-in connector
4.
Figure 105: Loosen screw connections
Loosen the union nut of the plug-in
connector on the left side by turning it
anticlockwise and unplug the connector.
Loosen the eight screws.
ATTENTION!
The measuring head may fall down.
Risk of impact and crushing.
➔
Hold the measuring head firmly
with one hand.
5.
Figure 106: Remove the measuring head
6.
Figure 107: Place the mounting plate seal in
position
Remove the measuring head and send it
back to Schaller Automation. Remove the
flat seal. Dispose of the seal in
accordance
with
local
disposal
regulations.
Place the new seal (supplied) on the
back of the new measuring head. Secure
the seal by inserting two of the eight pre-
assembled screws through the four outer
holes.