background image

IPB50 

10 

 

SELECTABLE OPERATING 

 

In addition to the STANDARD weighing Fn, with the basic abilities of TARE deduction and transmission of data 

(to printer, PC, or remote unit), the IPB50 series indicators can perform one of the following selectable 

functions: SIMPLE DISPLAY, HOLD, PEAK, ACCUMULATION TOTALIZER, BATCHING TOTALIZER, PIECE 
COUNTING, STANDARD, NET/GROSS, GROSS WEIGHT SET POINT, NET WEIGHT SET POINT. 

The user can select these operating. 

Refer to “USER SET-UP”. 

 

 

UISS

 (the tare value is lost when the IPB is switched off)

 

Press the key ->T<- to display the net Value 0 (led NET is turned ON) 

 

HOLD  (HLd) 

Operating Fn similar to standard Fn, but when Fn is pressed, the value of the weight is held on the display, the 

"Hold" LED turns on, and the display shows HoLd alternately with the weight held value (every 5 sec). To 
release the weight value on the display, press Fn key again.  

 

PEAK  (PlC) 

Operating Fn similar to standard Fn, but when Fn is pressed, the value of the maximum measured weight is 

held on the display, the "PIC" LED is on, and the display shows “-PlC-“ (PEAK) alternating with the held weight 
value (every 5 sec). This feature is well suited for “break load” testing.  Press Fn to enable the peak capturing 

function: the "PEAK" LED will turn on and the display will show the maximum detected weight, alternated with 

the writing PIC.  To stop the test and return to standard operation press Fn key again.  The weight value will 
be the maximum, before a fast decrease of the weight (i.e. measurement of the breaking strength). It's 

possible to set up the minimum time of the detectable weight peak.  To program this threshold time, press the 

 key for 5 sec. “- tP-“ will appear on the display and a number corresponding to the minimum peak time in 

0.01 seconds units.  By pressing the ->T<- key the user can scroll time values 2, 5, 10, 20, 50,100, and 200. 
Press 

 in order to confirm the selected value. The normal used value is 2 (fastest capture).  

Nota : Select the value of the filter (Mode) at LO-1 

 

ACCUMULATION TOTALIZER (tot 0) 

 

In this operating, it is possible to sum sequential weighments. When Fn is pressed, the displayed weight 

accumulates into a total weight in the indicator’s memory.  In order to avoid repeating a particular weighments 
the Fn key activates only according to the “rIAt” parameter setting in the USER SETUP (determined by zero 

weight or instability)..  If a printer is connected, the Fn key commands the printing of the weight value.  After 

initiating a Horizontal Totalizer, pressing Fn also allows the user to temporarily view the number of weighments 
and the accumulated TOTAL weight at the moment (SUbtotal). If the length of the numeric value (of the 

accumulated total weight) is more than 5 digits the displayed weight value will occur in two steps.  
To cancel the accumulated TOTAL press 

: If a printer is connected, before zeroing, the total NET weight (or 

subtotal) is printed along with the number of weighments.  When the display shows "ZERO", the Fn key is not 

enabled for totalization, only the accumulated total is displayed.  
 

BATCH TOTALIZER (tot S) 

This Fn prepares a sum for “recipe-batching.” Similar to the previous accumulation totalizer, this totalizer 

proceeds by allowing the weight to accumulate on the scale without unloading the previous “ingredient”.  When 

Fn is pressed, the indicated weight is added to the subtotal, and the displayed weight returns to a value of 
zero.  

 

 

 

Содержание IPB 50

Страница 1: ...WEIGHING INDICATOR IPB 50 Available for version 1 05 or more INSTRUCTION MANUAL SCAIME SAS tel 33 0 4 50 87 78 64 BP501 fax 33 0 4 50 87 78 46 74105 ANNEMASSE FRANCE WEB SITE www scaime com ...

Страница 2: ...FICATIONS page 3 POWER SUPPLY START UP page 4 FRONT PANEL KEYS AND INDICATORS page 5 IPB50 BASIC FUNCTIONS page 6 USER SET UP page 7 TECHNICAL CONFIGURATION page 15 SERIAL OUTPUTS page 21 CONNECTIONS page 25 EXAMPLE OF IPB50 PRINTING page 29 ...

Страница 3: ... the data through its two serial ports WARNING Any attempt to repair or alter the unit can expose the user to the danger of electric shock and it will void our warranty This instrument is covered under warranty provided that THE USER HAS NOT for any reasons OPENED IT If any problem with the unit or system has been experienced please notify the manufacturer or the dealer from which the instrument w...

Страница 4: ...IPB50 3 SPECIFICATIONS ...

Страница 5: ...o the same outlet Don not step or squash the mains cable The instrument will first show a random number then the code X Y ZZ where X indicates the keyboard type X 0 means 4 key keyboard and Y ZZ indicates the installed software version Then capacity and minimum division are shown The instrument will perform an auto test and warm up routine After few seconds the instrument will indicate ZERO if the...

Страница 6: ... the unit of measure in use lb or setpoint 1 is on 6 If turned on indicates the functional status of the instrument peak hold and piece counting operations or setpoint 2 is on Performs an automatic ZERO when the signalled weight is within 2 of the total capacity LEDà0ßis on Cancels a negative tare value resulting from a previous tare with the load now removed The displayed value of the weight pres...

Страница 7: ...shed it may be any number between 1 and 65535 which is kept in memory even when the instrument is turned off To zero the TICKET NO one must go to the ntlk parameter in USER SET UP and select rEStl and push The TICKET NO is printed before the date hour and may be used in all functioning Fns and in the totalizer Fn it increases only with the resetting of the total PRINT OF HEADING In order to enter ...

Страница 8: ...or with Net Gross conversion Fn UISS Weight indicator HLd Weight indicator HOLD function holds last measured weight PIC Weight indicator PEAK detection function shows highest recorded weight tot 0 Weight indicator totalizer functionaccumulation tot S Weight indicator totalizer batch COUn Weight indicator counting function StPG Weight indicator setpoint for gross weight StPn1 Weight indicator setpo...

Страница 9: ... AOZE starts to blink With this step it is possible to adjust the output value of the analog interface when the scale displays a zero weight The function of the keys are the following 0 forces the analog output at the minimum value T increases the output value of the analog interface in small increments Fn increases the output value of the analog interface in a large increments fine adjustment and...

Страница 10: ...ts date time for every single totalizer and in the total weight ntlk tlk 0 Doesn t print the progressive number on the ticket tlk 1 Prints the progressive number on the ticket rEStl By pressing the ticket s progressive number is reset rlAt ZerO the totalization and the printing are reactivated with the return of zero of the net weight InSt the totalization and the printing are reactivated with the...

Страница 11: ...To program this threshold time press the key for 5 sec tP will appear on the display and a number corresponding to the minimum peak time in 0 01 seconds units By pressing the T key the user can scroll time values 2 5 10 20 50 100 and 200 Press in order to confirm the selected value The normal used value is 2 fastest capture Nota Select the value of the filter Mode at LO 1 ACCUMULATION TOTALIZER to...

Страница 12: ...OTE After the PCS counting step has been completed and an APW value has been stored in memory by pressing Fn once the WEIGHT of the pieces will be displayed Pressing Fn a second time will display APW To see the APW value that the instrument has recorded press two times The first time that is pressed the first 6 digits of the unit weight value will be displayed only the digits before the decimal pl...

Страница 13: ...a Their relative LED displays the appropiate unit Only this Fn has the ability to convert from kg to lbs and back again However the other Fns may be chosen in either unit of measure during configuration beforehand NET GROSS NtgS the tare value is lost when the IPB is switched off Press the key T to display the net Value 0 led NET is ON however with the Fn key having a net gross switch If there is ...

Страница 14: ... of the SETPOINTS until a new value is confirmed To set the initial status of the two relays normally open N O or normally closed N C refer to the USER SET UP ANALOG OUTPUT optional In all function if the optional analog output is present it is possible to use the 0 10V FIXED output and the 0 20 mA CONFIGURABLE output jumper on the OPEN analog board or in 4 20 mA closed JUMPER The voltage and the ...

Страница 15: ...ard 0 20 4 20 mA of the analog output refer to terminal board connections on the IPB50 s back Through the AOMA AOZE and AOMI steps adjust the output voltage value which one wants in respect to the f s capacity with zero weight and the minimum value of the analog output NOTE IT IS NECESSARY FOR THE INSTRUMENT TO BE CALIBRATED ...

Страница 16: ...ease a numerical value of a digit selected with Fn key Fn Changes the numerical value of a step by selecting the digit to be changed it moves forward backward the selected digit The digit selected is the flashing one Use T to increase the value The display shows each step for which below you can find an explanation The symbol shows the configuration normally set by the manufacturer relative to tha...

Страница 17: ...ED shows dECI then press set the decimal point position and go ahead with the configurations step 0n 0 AUTO 0 AT SWITCH ON Enables AUtO or disables FrEE the ability of the instrument to acquire a gross zero automatically when the instrument is first switched on When AUtO is enabled the display will be zeroing if the gross weight value on the scale is less than 10 of the maximum capacity AUtO tIM0 ...

Страница 18: ...ter the actual value of the reference weight that will be used for the 1st calibration point Use the Fn and T keys to select and confirm with tP I flashing 1ST POINT CALIBRATION Place a reference weight on the platform of the same value as the one programmed in the previous step ddtI Wait for the scale to stabilize then confirm the calibration by pressing tP1 stop flashing wait a few seconds for t...

Страница 19: ...ansmission of the weight on the one way serial output in the standard format of the transmission protocol useful for connecting to PC when bi directional serial output is already in use Pr n0 disables printer SET IF THE 485 TRANSMISSION IS USED Pr Lt1 enables printer to use normal height characters Pr Lt2 enables printer to print date time and normal height characters Pr Ht1 enables printer to pri...

Страница 20: ...he input connection By doing so you emulate a key function by acting on the relative input your choices are as follows nOnE no key ZErO 0 key TArE T key ModE Fn key EntEr key StArt lock the keys Input is good only when OPT GND contact goes from open to close nOnE L0Ad flashing DATA STORAGE AND CONFIGURATION MENU EXIT Once all the configuration operations have been set or modified in order to save ...

Страница 21: ...action between the calibration place and the place of use During calibration the values for the calibration zone are programmed in a specific step If the instrument has to work in a different geographic zone than it s original one it s necessary to modify it by configuring the relative parameter This avoids error in weight values from different forces of gravitational attraction NOTE TABLE 1 below...

Страница 22: ...put the communication can take place at greater distances greater than 100 m than with the RS232 in this case the communication protocol is the same as that of the RS232 without identifying code RS485 TRANSMISSION PCMO 485 This serial transmission it is used to connect more than one IPB50 to a PLC or PC The indicator will ask to insert a two digit identification code from 00 to 99 The data transmi...

Страница 23: ...DE 485 Read command READ CR LF CCREAD CR LF Answer see below TARE command TARE CR LF CCTARE CR LF Answer ACK ASCII 06H if command is understood Nothing or NACK ASCII 15H if command is not understood Zero command ZERO CR LF CCZERO CR LF Answer ACK ASCII 06H if command is understood Nothing or NACK if command is not understood Relay 1 Enabling Setpoint STP1XXXXX CR LF CCSTP1XXXXX CR LF Answer ACK AS...

Страница 24: ...t is in OVERLOAD when the gross weight is over 6 009 kg 6 000 0 009 The scale displays UNDERLOAD _ _ _ _ when there is a negative gross weight on the scale In the CONTINUOUS transmission Fn through EXTERNAL COMMAND or WITH EVERY WEIGHING the string transmitted from the indicator is relative to the weight displayed GROSS WEIGHT DISPLAYED TRANSMITTED STRING UNDERLOAD UL GS 0000000 uu CR LF where 000...

Страница 25: ...5 RELAY 2 6 COMMON RELAY 7 RELAY 1 POWER SUPPLY 8 GND O Vdc 9 12 to 25 Vdc J1 INPUT FOR IPB50 ON BUTTON J2 IF CLOSED PRINTER CONNECTION WITH RS232 J3 IF CLOSED PRINTER CONNECTION WITH TTL J4 INPUT FOR TILT SENSOR Nota All the connection GND are connected together SENSE AND EXCITATION SENSE AND EXCITATION must be connected together with 4 wires sensor MORSETTIERA A MORSETTIERA B 1 2 3 4 5 6 7 8 9 1...

Страница 26: ...ARD CONNECTION J1 EEPROM INITIALISATION AND ACTIVATION OF DEFAULT PARAMETERS J2 IF CLOSED IT ENABLES IN CONFIGURATION THE ACCESS TO METHROLOGICAL PARAMETERS J3 AC CLOSED EPROM MEMORY BC CLOSED MEMORY FLASH J5 NO FUNCTION ...

Страница 27: ...ITATION 10 GND O Vdc 11 12 25 Vdc ANALOG OUTPUT JUMPER J1 OPEN 0 20 mA CLOSED 4 20 mA Nota All the connection GND are connected together SENSE AND EXCITATION SENSE AND EXCITATION must be connected together with 4 wires sensor Max load for analog output 0 20 or 4 20mA 500 OHM You must not have the and input reading system connected at the GND risk of destruction of the current output 1 2 3 4 5 6 7 ...

Страница 28: ...CKET NR 00002 12 03 08 01 02 Totalizer Fn With Register tot o tot s JOHN BLACK BRISTOL CRAVEN ROAD N 5 TEL 555 24773211 GROSS 5 124 kg PRE TARE 1 350 kg NET 3 774 kg TICKET NR 00003 12 41 08 01 02 Indicator in Standards Fn UISS Std ntgS GROSS 3 330 kg PCS 1009 APW 0 00330 kg ID1 1326 ID2 954 12 11 08 01 02 Piece Counting Fn COUn GROSS PEAK 0 726 kg ID1 5931 ID2 9863 12 09 08 01 02 Peak Fn PIC GROS...

Отзывы: