IPB50
14
4 / 20 mA
MAKE SURE THAT THE JUMPER, ON THE ANALOGIC BOARD IS CLOSED.
Connect the digital multimeter on the terminal board 0 / 20 , 4 / 20 mA of the analog output (refer to terminal
board connections on the IPB50’s back).
Through the AOMA, AOZE, and AOMI steps, adjust the output voltage value which one wants in respect to the
f.s. capacity, with zero weight and the minimum value of the analog output.
NOTE
: IT IS NECESSARY FOR THE INSTRUMENT TO BE CALIBRATED.