IPB50
18
UnIt
(*) MEASURE UNIT SELECTION
Allows user to select which unit of measure will be printed:
g / kg / t / Lb.
(!)
kg
M0dE
FILTERING – DISPLAYING
The 9 possible programmable values are the following:
HI-
1
HI-
2
HI-
3
= maximum immunity to vibrations, the display update slower, but with longer gaps.
HS-
1
HS-
2
HS-
3
= medium filter, continuous variations of the weight.
L0-
1
L0-
2
L0-
3
= light filter the displayed value changes continuously - not recommended in weighing
systems with over 3000 divisions of resolution.
The numbers
1, 2, 3
after the filter type indicate its effectiveness from weakest (1) to highest (3).
DInA
DYNAMOMETER
Applications such as crane scales have particular requirements for obtaining stability and countering the
tendency to change the value of weight displayed.
“DInA
0”: Dynamometer operating Fn disabled.
“DinA
1”: Dynamometer operation Fn enabled.
(!)
0
PrIn
PRINTER SETUP
This feature enables or disables the printer
ALL
= continuous transmission of the weight on the one-way serial output in the standard format of the
transmission protocol (useful for connecting to PC when bi-directional serial output is already in use).
Pr - n0 = disables printer.
(SET IF THE 485 TRANSMISSION IS USED)
Pr – Lt1 = enables printer to use normal height characters.
Pr – Lt2 = enables printer to print date/time and normal height characters.
Pr – Ht1 = enables printer to print double height characters.
Pr - Ht2 = enables printer to print date/time and double height characters.
(!)
Pr-no
PcMO
DATA TRANSMISSION
This step is used to select the data transmission on the BI-DIRECTIONAL serial port.
OndE
= transmission on external command (PC).
rIPE4
= transmission to 4 digit remote display.
rIPE6
= transmission to 6 digit remote display
.
485
= serial transmission
with communication protocol 485. If selected the indicator will prompt
you
to insert the identification code (between 00 and 99).
ALL
= continuous transmission.
StAb
= transmission at each weighing operation..
(!)
OndE
bAud
SET BAUD RATE
Use this step to select one of the following Baud rates for the BI-DIRECTIONAL serial port:
150/300/600/1200/2400/4800/9600 Baud.
(!)
9600