- 22 -
* The types (extensions) of files stored are as follows:
・
Main recording management file (dm)
This file contains the main recording start and end times, and link information.
A single file manages 50 dmt files.
・
Main recording trend file (dmt)
This file contains measurement data where the main recording has been split according to
the recording interval.
・
Main recording event file (dme)
This file contains event logs such as alarms and messages.
・
Sub-recording management file (ds)
This file contains the sub-recording start and end times, and link information.
・
Sub-recording trend file (dst)
This file contains measurement data where the sub-recording has been split according to
the recording interval.
③
Etc: Contains text tables (txt) that can be accessed by using the listed keys in the character
input screen, as well as parameter settings files (dps).
④
Prm: Contains settings files (dps).
These files can be saved and read by this equipment. Settings data created with the
software provided (Parameter Loader) must be stored in this folder.
* Data analysis
Installing the analysis software (Data Viewer) from the CD-ROM onto the PC enables data
analysis to be performed. For details on how to use this software, refer to the instruction
manual stored on the CD-ROM.
Data saved in CSV format can be edited using commercially available spreadsheet software
such as Excel.
Maintenance
● Replacing the wick
Caution
・
Before using this equipment, be sure to check if dust or dirt is not adhering to the wick.
If it is dirty, replace it to new one.
・
Regularly check the water level in the water reservoir and check if the water is clean.
If the water level is too low or the water is dirty, measurement errors may occur.
It is recommended to replace the wick and water, and clean the water supply pot once a
week.
・
When handling the wick, use a pair of tweezers or clean your hands with soap to prevent the
transfer of dirt (such as grease, salt or acid). Adhering dirt such as grease or salt may result
in measurement errors.