background image

 

- 2 - 

select a brushless motor when placing your order. The test results under the normal 

temperature and humidity conditions demonstrate that the service life of brushless motors is 

at least 10,000 hours of continuous operation. 

  Do not operate the touch panel with a pointed object such as a pen. Doing so may scratch 

the screen.

 

 

  Caution 

■ Be sure to observe the following precautions in order to use this equipment correctly. 

  Be careful to correctly wire the equipment. Otherwise, the equipment may fail. 

  We recommend the use of protective cover to prevent accidental contact with the wiring part.   

  Be sure to turn off the power supply of the equipment before starting any wiring work to prevent 

electrical shock.   

(1) To prevent electrical shock, be sure to perform  protective grounding before turning on 

the power to the equipment. 

(2) Do not cut or disconnect the protective grounding wire. 

(3) Do not turn on the equipment before connecting the wires or setting up the sensor. 

 

Make sure that the power is off when connecting the wires. 

  Make sure that the power source conforms to the power supply voltage of this equipment. 

(1) Power supply rating 

: 100 to 240 VAC, 50/60 Hz, 25 VA max. 

(2) Operating voltage range  : 85 to 264 VAC 

Note:  The  power  cord  provided  is  exclusively  used  for  100  VAC.  If  the  equipment is 

used with a different voltage, prepare a cable that matches the voltage and the 

shape of the power outlet. 

  Do not use unconnected terminals for any other purposes such as relays. 

  We recommend that the protective grounding wire is not shared with other devices.   

      If shared, noise from the grounding wire may affect this equipment. 

 Do not operate this equipment in an area subject to flammable gas, explosive gas, corrosive gas, 

spattering water, or steam. 

    Installation location 

: Indoors 

    Altitude 

: No higher than 2,000 meters above sea level 

    Installation ambient 

: 0 to 50ºC, 20 to 80% RH (no condensation) 

    Overvoltage category : Category II 

    Permissible exposure limits for contaminants 

                               

: Level 2 

    Vibration 

: 10 to 60 Hz, 0.2 m/s

2

 

    Impact 

: Not allowed 

  This equipment has a built-in battery for a clock.    If the battery replacement is required, contact 

the store where you purchased this equipment or our service network by all means. 

  Do  not  carry  or  move  the  equipment  with  the  sensor  and  the  paperless  recorder  connected. 

Doing so may cause them to fall, resulting in a very dangerous situation. 

Содержание 7435-50

Страница 1: ...Instruction Manual No 7435 50 Multiple Measuring System Model SK 5RAD MR 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english P 12...

Страница 2: ...ipment in an area subject to dust flammable gas explosive gas corrosive gas SO2 H2S etc etc Using this equipment in an environment with electrical noise may produce an unstable display or an increase...

Страница 3: ...supply voltage of this equipment 1 Power supply rating 100 to 240 VAC 50 60 Hz 25 VA max 2 Operating voltage range 85 to 264 VAC Note The power cord provided is exclusively used for 100 VAC If the equ...

Страница 4: ...ames product names system names etc are trademarks or registered trademarks of their respective companies Overview The operating principle of this equipment is based on the wet and dry bulb system whi...

Страница 5: ...s must specify this option when placing an order Components Sensor Hanging ring Used to suspend the sensor Wall mounting bracket Used to mount the sensor on a wall Power switch Turns the DC motor on a...

Страница 6: ...toward you while pushing the two knobs downward To close the cover push it straight back in place while holding down the knobs Terminal block Used to connect the power cord and sensor cable LAN conne...

Страница 7: ...11 12 13 14 15 16 17 18 19 Dry bulb temperature CH1 Wet bulb temperature CH2 A B B A B B Wire color R terminal color Red Red White Red Black Red Orange Blue Blue Blue Brown Blue Note For brushless mot...

Страница 8: ...en installing the sensor on a wall use the appropriate screws for the installing location The sensor can also be hanged by using the hook on the sensor head A tripod for hanging the sensor is optional...

Страница 9: ...ile phones Level places forward tilt 0 backward tilt between 0 and 30 Water supply method Caution When handling the string of wick use a pair of tweezers or wash your hands with soap to prevent the tr...

Страница 10: ...quipment was purchased or our service network If measured values are abnormal such as measured value exceeds the measuring range or the dry bulb temperature is lower than the wet bulb temperature B OU...

Страница 11: ...d the SD memory mark is lit in red in the memory status display section During this state never remove the card Doing so may damage the data and cause the equipment to fail You can freely change the r...

Страница 12: ...OM provided Installation Turn off any active applications currently running before installing this software If the previous version has been installed delete the software using Add Remove Programs in...

Страница 13: ...s optional functions however these cannot be used The Engineering key on the menu screen is used for the factory adjustment only Do not change the settings for these functions Note that any loss of or...

Страница 14: ...d In this state never remove the SD memory card MiniSD and MicroSD memory cards cannot be used If an adapter for MiniSD or MicroSD is used for these cards there is a risk of not being able to remove t...

Страница 15: ...ouching this area enables the screen title to be changed Time display This area displays the current date and time Trend changeover This area is used to switch the type of trend screen REAL or HIST No...

Страница 16: ...D memory card is mounted or an SD memory card that is mounted can be removed Light blue An SD memory card has been mounted Red The equipment is communicating with the SD memory card Data is being writ...

Страница 17: ...ry overwriting OFF Data is not overwritten when the SD memory card becomes full Storage format CSV data and binary data Recording type Instantaneous value Graph display Digital FUNC button Screen swit...

Страница 18: ...CH13 and dew point temperature CH14 Setting Main Touch this button to set up recording Items such as the recording interval can be set here LCD Touch this button to set the screen brightness and auto...

Страница 19: ...lly created when an SD memory card is mounted on this equipment To load the stored settings into the equipment touch Param load Selecting the desired settings from all the stored settings loads the se...

Страница 20: ...e Set the interval for recording measurements to a file The selectable range varies according to the specified value in Record Cycle File overwrite Select to overwrite data or not when the SD memory c...

Страница 21: ...interval is one second the maximum and minimum values of 10 values measured at intervals of 100 ms will be recorded OFF Trend data only is recorded on the Trend screen and is not recorded in the SD m...

Страница 22: ...erting an SD Memory Card into this equipment automatically creates a folder named Recorder where four subfolders will be created The subfolders are as follows Cap Contains image files bmp that are cap...

Страница 23: ...saved and read by this equipment Settings data created with the software provided Parameter Loader must be stored in this folder Data analysis Installing the analysis software Data Viewer from the CD...

Страница 24: ...e clamp convex part is placed and aligned with gloves of the upper and lower sections of the air intake tube At that time don t let the string get stuck or caught between the housing and air intake tu...

Страница 25: ...al strands warps Make a string by binding the remaining wraps of gauze 8 Put back the lower part of air intake tube clamp and insert the string into the water reservoir Follow steps 4 5 6 in Replacing...

Страница 26: ...s of operation If the motor needs to be replaced due to aging contact the store from which the equipment was purchased or our service network For the continuous operation for a long period of time sel...

Страница 27: ...y SD memory card SD max 2GB SDHC max 32GB Water resistance JIS C 0920 IP65 Only for the front of the display with the use of packing for the panel mount Power Requirements 100 to 240VAC 10 50 60Hz 25V...

Страница 28: ...customer misuse Out of warranty products will be repaired on charge basis Warranty during transportation The warranty during transportation only applies to the products that we arrange the shipment Th...

Отзывы: