- 8 -
Water
supply
bottle
Fill water
four-fifths
full
Water
reservoir
②
Installing the paperless recorder
Install the paperless recorder in the places as following.
・
Places free from mechanical vibrations or impact
(Vibration: 10 to 60 Hz, 0.2 m/s
2
, impact: not allowed)
・
At dust-free atmosphere
・
Places free from flammable gas, explosive gas and corrosive gas (such as SO
2
or H
2
S)
・
Places within an ambient temperature range of 0 to 50ºC and with small temperature variation.
・
Places away from direct high radiant heat and direct sunshine.
・
Places free from water droplets and condensation within a humidity range of 20 to 80%rh.
・
Well-ventilated places to diffuse heat from the equipment
・
Places away from electromagnetic waves from wireless equipment or mobile phones
・
Level places (forward tilt: 0°, backward tilt: between 0° and 30°)
Water supply method
Caution
・
When handling the string of wick, use a pair of tweezers or wash your hands with soap to
prevent the transfer of dirt (such as grease, salt or acid).
・
If the wick or the sensor section becomes dirty, accurate measurements cannot be made.
Use distilled water for the water supply whenever possible.
・
Do not wet the inner wall of the air intake tube. Otherwise, the wet bulb temperature will
become unstable, which may result in measurement errors.
1.
Carefully pull the string from the guide above the water
reservoir and put the string into water reservoir.
2. Fill the water reservoir to four-fifths with distilled water,
using the water bottle supplied (Approx. 13 cc).
Pour water to the string so that the temperature bulb
also gets wet.
As for the replacing the wet bulb wick, refer to the
“Maintenance” on the page 22.
Air intake clamp
String
Water reservoir