7.15.4 Drying (F4)
After cleaning, drying follows. The drying program is structured analogously to the
cleaning program. Here too you will be interactively guided through the individual
steps. The diagram will show you how the connections should be produced.
Description of the connections:
a:
IN – Inlet pressure tubing
Connect inlet pressure tubing
b:
Venting 2
Connect this outlet as defined for each individual step
c:
OUT – Outlet pressure tubing to the filter housing
Connect this outlet as defined for each individual step
d:
Venting 1
Connect this outlet as defined for each individual step
e:
Ext.Ref.Tank
Connect this outlet as defined for each individual step.
You will need a coded Stäubli connector, see “Accessories.”
f:
Receptacle – Fluid holding tank
For collecting the fluid still remaining in the pneumatic block and internal
reference tank
Laboratory peristaltic pump or diaphragm pump
Power max: 1000 ml/min
During the drying cycle two (2) compressed air sources are required
(connections a & e)
7.15.5 Pressure Gauge Function (F5)
With the pressure gauge function, you can use the internal pressure sensor for
gas pressure measurement. For measurement with the internal pressure
sensor, the sensor (pressure gauge) is connected to the outlet. The remaining
pneumatic components are turned off. Now, Sartocheck
®
3 plus works like a digital
pressure gauge of accuracy class 0.1 and can be used to calibrate other pressure
gauges. During measurement with an external pressure sensor, Sartocheck
®
3 plus
operates like a digital pressure gauge of accuracy class 0.1 and can also be used
to calibrate other pressure gauges.
Do not use the pressure gauge function to monitor the pressure during
liquid filtration!
7.15.6 Changing Own Password (F6)
According to the settings for the password expiration date, the user must change
his or her password. By entering an old password, the user can store a new
password and confirm. This password will then be valid until the defined expiration
date is reached.
For more details, please refer to the menu:
Service -> Software configuration -> Management unit parameters -> Parameter
processing-> Password