Cubis MSA User Manual
17
Getting Started
Installation Instructions
t
Select the proper setup location:
– Set up the device on a stable, even surface that is not exposed to vibrations
(e.g. weighing stone).
– Place the device in a location with enough free space around it so that
excessive heat cannot build up.
– Maintain free access to the device at all times.
Choose a location that is not subject to the following negative influences:
– Heat (heater or direct sunlight)
– Drafts from open windows, AC systems and doors
– Extreme vibrations during weighing
– Excessive moisture
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is
brought into a substantially warmer area.
To avoid the effects of condensation, condition the balance for about two hours at
room temperature, leaving it unplugged from the power supply.
Carefully read all warnings and safety precautions in the respective section of this
manual.
Setting the Angle for the Display and Control Unit
The angle of the display and control unit can be adjusted by the user, in order to
ensure optimal readability of the weight values on the display at all times.
t
On MSA and MSU models, the display and control unit can be tilted as desired.
Power Cord Receptacle
t
Check the voltage rating and plug design.
– Make sure that the voltage rating printed on the manufacturer’s ID label is
identical to that of your local line voltage.
– If the stated supply voltage or the plug design of the power cord does not
comply with your country’s standard, please inform the nearest Sartorius
representative or your dealer.
– The power connection must be made in accordance with the regulations
applicable in your country.
– In order to connect the equipment to the power supply (protection class 1),
use a correctly installed wall outlet with a protective grounding conductor (PE)
and a fuse of maximum 16 A.
– If a main feeder from the ceiling is required or if installation of a CEE plug is
necessary, these must be carried out by a qualified electrician.
– Only use original Sartorius AC adapters.
°C
Содержание MSA3.6P
Страница 184: ...184 Cubis MSA User Manual Balance Dimensions Semi microbalances All dimensions are given in millimeters...
Страница 192: ...192 Cubis MSA User Manual Declarations of Conformity...
Страница 193: ...Cubis MSA User Manual 193 Declarations of Conformity...
Страница 194: ...194 Cubis MSA User Manual Declarations of Conformity...
Страница 195: ...Cubis MSA User Manual 195 Declarations of Conformity...
Страница 196: ...196 Cubis MSA User Manual Declarations of Conformity...
Страница 197: ...Cubis MSA User Manual 197 Plates and Markings...
Страница 198: ...198 Cubis MSA User Manual Plates and Markings...
Страница 199: ...Cubis MSA User Manual 199 Plates and Markings...
Страница 200: ...200 Cubis MSA User Manual Plates and Markings...