Cubis MSA User Manual
87
Calculation
t
Enter a name for the event print e. g. “Area” (max. 6 characters).
t
Define how many decimal points should be used for the results:
PQPGFKIKVU
t
Enter the unit to be used for the results
(max. six characters, e.g. cm
2
)
y
An overview of all settings is displayed.
t
Check all settings and change as required.
y
A message asks whether you would like to add another application.
y
Prompts for configuring weighing and printing appear.
t
Check all settings and change as required.
t
Enter a short name and a description for the new task.
t
To save the new task, touch
5CXG
.