Cubis MSA User Manual
27
Getting Started
2.
Push the cover of the below-cell weighing port to one side.
3.
Carefully screw on the below-cell weighing hook.
3
Do not screw it in too tightly,
as this could damage the thread or the balance.
h
Install a draft protection shield.
t
Attach the sample (e.g., using a suspension wire) to the hook.
4
When weighing has been completed, unscrew the hook and return it to the clip.
t
Close the cover of the below-cell weighing port.
The below-cell weighing port may not be opened or used on balances
used for legal metrology.
Balances with a readability of
)
1 μg
t
Remove both screws from beneath the weigh cell and detach the cover plate.
t
Attach a wire, for example, to the sample and hang it on the notched hook.
h
Install a draft protection shield.
2
3
4