Cubis MSA User Manual
63
Task Management
Example: Automatic printout to peripherals port; remote display operating mode
y
At a pre-set interval, automatic printouts should be sent to the ported
peripheral device as remote display.
Interface: Com A (peripherals port)
Protocol: Remote display (toggling to the interface configuration possible)
SBI Printout: Automatic printout
Criterion: Without stability, at stability or after load change
Format: 22 characters (for applications)
Interval for automatic printing: Display updates (1, 2, 5, 10 to100) or every 0.1, 0.2,
0.5 to 10 seconds.
Example: Printout to peripherals port only by using the PRINT
P
key; printer operating
mode
y
After pressing the Print key, the measured value should be printed to the ported
peripheral device.
Interface: Com A (peripherals port)
Protocol: Print (toggling to the interface configuration possible)
Standard data output: Standard 1 (measured value only), Standard 2 (measured
value and date/time), Standard 3 (measured value with date/time and weight block)
GLP-compliant printout Turned on or off
Print event: Press the PRINT key
Elements for the Print event PRINT: Measured value PRINT.MEASURE
Tare after printing: Turned on or off
Setting Options of Printout Parameters
Standard Data Output
You can select which data should be printed out as default.
5VCPFCTF
(measured value only)
5VCPFCTF
(measured value with date/time)
5VCPFCTF
(measured value with date/time and weight block)
7UGTKPRWV
is only selectable if the standard settings have been modified by
the user.
GLP-compliant Printout
You can select if and when the printout should be GLP-compliant
1HHPQ).2RTKPVQWV
(QTECNKDTCVKQPCFLWUVOGPVQPN[
#EVKXCVGFKPVJGCRRNKECVKQP
#NYC[UQP
7UGTKPRWV
is only selectable as long as none of the three options apply:
i.e. if only a GLP header but not a GLP footer has been configured.
Содержание MSA3.6P
Страница 184: ...184 Cubis MSA User Manual Balance Dimensions Semi microbalances All dimensions are given in millimeters...
Страница 192: ...192 Cubis MSA User Manual Declarations of Conformity...
Страница 193: ...Cubis MSA User Manual 193 Declarations of Conformity...
Страница 194: ...194 Cubis MSA User Manual Declarations of Conformity...
Страница 195: ...Cubis MSA User Manual 195 Declarations of Conformity...
Страница 196: ...196 Cubis MSA User Manual Declarations of Conformity...
Страница 197: ...Cubis MSA User Manual 197 Plates and Markings...
Страница 198: ...198 Cubis MSA User Manual Plates and Markings...
Страница 199: ...Cubis MSA User Manual 199 Plates and Markings...
Страница 200: ...200 Cubis MSA User Manual Plates and Markings...