Operating Instructions Combics CAIXS2
101
Data interfaces
Data Input Format
You can connect a computer to your scale to send commands controlling weighing
instrument functions and applications via the interface port.
All commands use the same data input format. They begin with the character
ESC
(ASCII: 27) and end with a carriage return
CR
(ASCII: 13) and
LF
(ASCII: 10). The
total length of a command is anywhere from 4 characters (1 command character
between the start and end described above) to a max. of 7 characters (4 command
characters). This number can also be higher when sending texts.
The commands listed in the following table must each be supplemented with
ESC
...
CR LF.
Example
: The command character for output is “P” (“output to Port”). To trigger this
command, send the string: “ESC P CR LF”.
Command Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Lock keys
P
Send display value to data interface
Q
Output acoustic signal
R
Release keys
T
Tare and zero
(combination tare function)
f3_
Zero (see also the “kZE_" command)
f4_
Tare without zeroing (see also “kT_" command)
i_
Information about the indicator, example of output: “C2/016202/1"
Explanation: Indicator: Combics 2, software version: 016202, active weighing platform: 1
kF1_
F1: Trigger
k
key function
kF2_
F2: Trigger
c
key function
kF3_
F3: Trigger
r
key function
kF4_
F4: Trigger
O
key function
kF5_
F5: Trigger
w
key function
kF6_
F6: Trigger
I
key function
kF7_
F7: Trigger
g
key function
kF8_
F8: Trigger
H
key function
kF9_
F8: Trigger
r
key function
kF10_
F8: Trigger
J
key function
kF11_
F8: Trigger
k
key function
kF12_
F8: Trigger
L
key function
kCF_
CF: Trigger
c
key function
kP_ Trigger
p
key function Print at printer interface
kT_ Trigger
T
key function (Tare)
kNW_ Trigger
n
key function (toggle the weighing platform)
kZE_ Trigger
(
key function (Zero)
x1_
Output model designation of active weighing platform, example: “LP6200S-0C"
x2_
Output serial number of active weighing platform, example: “0012345678"
x3_
Output software version of active weighing platform, example: “00-42-01"
x4_
Output software version of indicator, example: “01-62-01"
x9_
Output serial number of indicator, example: “0012345678"
x10_
Output model of indicator, example: “CAW2P4-1500RR-LCE"
z1_
Input: printout header 1
z2_
Input: printout header 2
txx...x_
xx...x: text input, length corresponds to input, for display in the measured value line
The ASCII code for the “underline” character (“_”) is 95 decimal.
Format for entering printout header lines: “ESC z x a ... a _ CR LF" with x=1 or 2
and a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF
characters.
Содержание Combics 2 CAIXS2
Страница 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Страница 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Страница 123: ...Operating Instructions Combics CAIXS2 123...
Страница 149: ...Operating Instructions Combics CAIXS2 149...
Страница 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Страница 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Страница 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 178: ...178 Operating Instructions Combics CAIXS2...
Страница 179: ...Operating Instructions Combics CAIXS2 179...