Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Configuring the A/D Converter (ADC)
The weighing platform must already be connected.
Opening the Menu Access Switch
The menu access switch is located on the back of the indicator, behind the cover.
t
Remove the cap.
t
Slide the switch to the right (= “open" position).
Menu Access Switch/Calibration Switch Cover
e
.
t
Switch off and restart the device.
.
.
t
While all segments are lit, press the
(
key briefly.
y
ADC-CON
appears briefly on the display, followed briefly by
S-CODE.
y
The cursor flashes on the display.
t
Enter the service password (see Appendix).
)
t
Confirm your entry using the
)
key.
y
The device is in Service mode. This can be recognized by the small
S
in the top
right of the display.
t
Select the weighing platform to be configured, using the
k
key to switch to
WP-2 if required.
)
t
Confirm your selection using the
)
key.
t
Select the Configuration mode using the
k
key:
STANDRD
or
VERIF
.
t
Carry out ADC configuration (see menu tree).
t
Once you have completed the configuration, save the data using the
SAVE
menu item.
y
The indicator will restart automatically.
Содержание Combics 2 CAIXS2
Страница 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Страница 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Страница 123: ...Operating Instructions Combics CAIXS2 123...
Страница 149: ...Operating Instructions Combics CAIXS2 149...
Страница 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Страница 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Страница 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 178: ...178 Operating Instructions Combics CAIXS2...
Страница 179: ...Operating Instructions Combics CAIXS2 179...