Operating Instructions Combics CAIXS2
53
Operation
Resolution
The resolution indicates the accuracy used to determine the reference weight.
The default setting is “display resolution." The resolution is increased when
“10-fold" or “100-fold" is selected. “10-fold" increases the resolution of the net
value by one step (display resolution x 10), “100-fold" increases it two steps
(display resolution x 100).
Setting:
APPLIC./APPLIC.1/COUNT./RESOLUT
menu item 3.9.
Parameter for saving weight values
The weight on the platform is saved as a reference value as soon as the platform
has stabilized. “Stability" is defined as the point at which the fluctuation of a
measured value lies within a defined tolerance. The narrower the tolerance,
the more stable the platform is at “stability".
The “increased stability" setting has a lower tolerance so that the average piece
weight saves is more accurate and the results more reproducible; however,
the measurement time can take longer.
Setting:
APPLIC./APPLIC.1/COUNT./SAVE WT.
menu item 3.11.
Reference sample updating
You can define whether or not the reference sample weight is updated
automatically during weighing using this setting. The reference sample weight
is updated automatically only when the following criteria are met:
1. “Automatic" must be set for reference sample updating in the menu.
2. The current piece count exceeds the original piece count by at least two.
3. The current piece count cannot be more than double the size of the original
piece count. This limitation does not apply to the first update if the piece
weight was entered via a keypad.
4. The current piece count is less than 1000.
5. The internally calculated piece count (such as 17.24 pcs) differs by less than
± 0.3 pcs from the nearest whole number (in this example: 17).
6. The weighing platform is stable in accordance with the parameter defined for
saving weights.
If automatic reference sample updating is selected in the menu and the piece count
(pcs) is displayed, the
Auto
symbol is displayed below the bar graph. If the reference
sample weight has been updated since you began weighing, the text line shows the
0pt
symbol. During an updating operation,
0pt
and the updated piece count are
displayed briefly in the measured value line. The new reference sample weight and
reference sample quantity are saved.
Setting:
APPLIC./APPLIC.1/COUNT./REF.UPDT
menu item 3.12.
Counting with two weighing platforms
You can use two weighing platforms simultaneously with the Counting application.
When using two platforms, you can choose from the following operating modes:
– Counting with two platforms of the same type
– Counting with one reference platform and one weighing platform.
Counting with Two Platforms of the Same Type
Use this mode to count different types of sample material with different weights.
For example, count the lighter-weight pieces on one platform and the heavier
pieces on another. You can define one of the two scales as the reference scale.
The reference scale is the first scale active when you switch on the device,
regardless of the setting for automatic initialization of the Counting application.
Counting with One Reference Platform and One Weighing Platform
In this operating mode, the reference platform is a high-resolution weighing
platform with a relatively low maximum capacity. The other platform is used for
weighing heavier samples, and has a high capacity with a relatively low resolution.
This allows you to both determine the reference sample weight with high
resolution; i.e., very precisely, and to count large amounts of parts, without
requiring an expensive high-resolution, high-capacity weighing platform.
Содержание Combics 2 CAIXS2
Страница 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Страница 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Страница 123: ...Operating Instructions Combics CAIXS2 123...
Страница 149: ...Operating Instructions Combics CAIXS2 149...
Страница 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Страница 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Страница 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Страница 178: ...178 Operating Instructions Combics CAIXS2...
Страница 179: ...Operating Instructions Combics CAIXS2 179...