Technical Data
15.10.2 Models BCE2202 | BCE1202
Model
BCE2202-1x (x = only CCN, OIN) | BCE2202i-1x
BCE2202-1x (x = only S, SAR, SJP, SKR)
BCE1202-1x (x = only CCN, OIN) | BCE1202i-1x
BCE1202-1x (x = only S, SAR, SJP, SKR)
Unit
Value
Value
Value
Value
Readability | Scale interval (d)
mg
10
10
10
10
Maximum capacity (Max)
g
2200
2200
1200
1200
Repeatability
At 5% load, typical value
mg
5
5
5
5
At approx . the maximum load, typical value
mg
10
10
10
10
Linearity deviation
Limits
± mg
20
20
20
20
Typical value
± mg
6
6
6
6
Sensitivity drift b10 °C and +30 °C
± ppm/K 2
3 .5
2
3 .5
Tare maximum capacity (subtractive)
<100% of maximum capacity
isoCAL (only for i-1x models):
Temperature change
K
2
2
Time interval
h
6
6
Only for models with approval:
Accuracy class
II
II
Type
BC_EE
BC_EE
Verification scale interval (e)
mg
100
100
Minimum load (Min)
mg
500
500
Minimum initial weighing according to USP (United States
Pharmacopeia), Chap . 41
Optimum minimum initial weighing
g
82
82
8 .2
8 .2
Typical measurement time
s
<
1 .0
<
1 .0
<
1 .0
<
1 .0
Typical stabilization time
s
<
0 .9
<
0 .9
<
0 .9
<
0 .9
Recommended Calibration Weight
External calibrated test weight
g
2000
2000
1000
1000
Accuracy class in accordance with OIML R111-1
F1
F1
F1
F1
Weighing pan size
mm
182 x 182
182 x 182
182 x 182
182 x 182
Net weight, approx .
kg
4 .3 | 6 .2
4 .3
4 .3 | 6 .2