36 Entris
®
BCE Operating Instructions
Technical Data
15.3.2 Power Supply Unit
Unit
Value
Type: Sartorius power supply unit YEPS01-15V0W
Primary
Voltage
V
AC
100 – 240 (
±
10 %)
Frequency
Hz
50 – 60
Current consumption, maximum
A
0 .2
Secondary
Voltage
V
DC
15 (
±
5 %)
Current, maximum
A
0 .53
Short-circuit protection
Electronic
Protection class according to IEC 60950-1
II
Pollution level according to IEC 61010-1
2
Overvoltage category according to IEC 60664-1
II
Other data: See label on the power supply unit
15.4 Electromagnetic Compatibility
Interference resistance:
Suitable for use in industrial areas
Transient emissions:
Class B
Suitable for use in residential areas and areas that are connected to a low voltage network that also supplies residential
buildings .
15.5 Backup Battery
Unit
Value
Lithium battery, type CR2032
Service life at room temperature, minimum
Years
10
15.6 Materials
Housing
Polybutylene terephthalate (PBT)
Control unit
Glass
Draft shield
Glass/polybutylene terephthalate (PBT)