22 Entris
®
BCE Operating Instructions
System Settings
Parameter
Setting values Explanation
LIGHT
OFF
Deactivates the lighting on the operating display .
ON*
Activates the lighting on the operating display .
* Factory setting
7.3.6 “DATA.OUT.”/”COM. SBI” Menu
Parameter
Setting values Explanations
COM . OUTP .
IND .NO*
Activates the manual data output without stability .
IND .AFTR
Activates the manual data output after stability .
AUTO .W/O
Activates the automatic data output without stability .
AUTO W/
Activates the automatic data output after stability .
STOP .AUT .
OFF*
Deactivates the option to abort the automatic data output .
ON
The automatic data output is aborted by pressing the [Print] button or a software command .
AUTO .CYCL .
EACH VAL*
Starts the automatic data output with a cycle after each value .
AFTER 2
Starts the automatic data output with a cycle after every 2nd value .
INTERV .
The output rate can be set from 0 - 9999 seconds under “INPUT/INTERV .” .
FORMAT
22 CHARS*
The data output provides 22 characters per line (16 characters for the measured value and
6 characters for identifiers) .
16 CHARS
The data output provides 16 characters per line for the measured value .
EXTR .LIN .
The data output provides an additional line with the date, time, and weight value .
AUTO .TAR .
OFF*
Deactivates automatic taring after data output .
ON
The device automatically tares after data output .
* Factory setting
7.3.7 “DATA.OUT.”/”PRNT.PAR.” Menu
Parameter
Setting values Explanation
ACTIVAT .
MAN . NO
Manual without stability: The print process can be started manually at any time .
MAN .AFTR*
Manual after stability: After pressing the [Print] button, the print command is only executed
once stability is achieved .
INTERV .
The output rate can be set from 0 - 9999 seconds under “INPUT/INTERV .” .
AUTO .LC
Automatically at load change: The print process starts after every load change .
FORMAT
22 CHARS*
The printer output prints 22 characters per line (16 characters for the measured values and
6 characters for identifiers)
EXTR .LIN .
The printer output prints an additional line with the date, time, and weight value .
PRT .INIT .
OFF
Deactivates the output of the application parameters .
ALL*
The print command prints all parameters .
MAIN .PAR .
The print command only prints the main parameters .
GLP
OFF*
Deactivates the GLP printout .
CAL ./ADJ .
Activates the GLP printout for all calibration and adjustment processes .
ALWAYS
The GLP printout is always switched on . All printouts contain a GLP header and a GLP footer .
TAR ./PRT .
OFF*
Deactivates automatic taring after printer output .
ON
Automatically tares the device after every printout .
* Factory setting