30 Entris
®
BCE Operating Instructions
Cleaning and Maintenance
8.12 Printing Weighing Result with ID Marking
The device, the sample and a batch can be assigned an ID
number . The ID numbers are exported during GLP-compliant
printing .
Requirements
− The identity number is specified (see Chapter “7 .3 .20
− The printing of the lot ID line in the GLP printout is
activated in the menu (see Chapter “7 .3 .20 “INPUT” Menu”,
page 26) .
− The “SPL . ID” output is activated in the menu .
− The GLP-compliant printout is activated (see Chapter “7 .3 .7
“DATA .OUT .”/”PRNT .PAR .” Menu”, page 22) .
Procedure
t
Start the printout . To do so, press the [Print] button .
y
The GLP header is printed with the ID marking of the
LOT ID set in the menu and the current weight value .
y
The [GLP] button appears in the operating display .
t
Confirm the [PRINT] button .
y
The SPL . ID and the current weight value are exported .
t
Confirm the [PRINT] button .
y
The SPL . ID and the current weight value are exported .
t
To exit the GLP printout: Press the [GLP] button .
y
The GLP footer is printed .
9 Cleaning and
Maintenance
9.1 Removing the Weighing Pan and
Associated Components
Requirements
− The device is switched off .
− The device is disconnected from the power supply .
Procedure
t
Remove the weighing pan,
the shock absorbers and
the frame draft shield .
9.2 Cleaning the Device
NOTICE
Corrosion or damage to the device due to unsuitable
cleaning agents!
t
Do
not
use corrosive, chloride-containing, or aggressive
cleaning agents .
t
Do
not
use cleaning agents that contain abrasive
ingredients, e .g . scouring agents, steel wool .
t
Do
not
use solvent-based cleaning agents .
t
Only use suitable cleaning agents (materials see Chapter
“15 .6 Materials”, page 36) and observe the product
information for the cleaning agent used .
Procedure
t
NOTICE
Malfunction or damage to the device due to the
ingress of moisture or dust!
t
Only slightly moisten cleaning materials, such as cloths .
t
Remove dust and powdery sample residue with a brush
or hand-held vacuum cleaner .
t
Wipe the device and the associated components with a
slightly damp cloth . Use a mild soapy solution or a suitable
cleaning agent for more severe contaminations .
9.3 Assembling the Weighing Pan and
Associated Components
Procedure
t
Assemble the weighing pan and the associated components
(see Chapter “5 .5 Assembling the Device”, page 15) .
t
Connect the device to the power supply (see Chapter “6 .2
Connecting the Power Supply”, page 16) .
9.4 Maintenance Schedule
Interval
Component Action
Regularly; depending on
the operating conditions
Device
Contact Sartorius
Service .
9.5 Software Update
For a software update, contact Sartorius Service .