Entris
®
BCE Operating Instructions
35
Technical Data
15 Technical Data
15.1 Ambient Conditions
Unit
Value
Installation site: For indoor use only, max . height above sea level
m
3000
Temperature
Environment (metrological data)*
°C
+10 – +30
Environment
°C
+5 – +40
Storage and shipping
°C
-10 – +60
Relative humidity**
At temperatures up to 31 °C,
non
-condensing
%
15 – 80
Then linear decrease from max . 80 % at 31 °C to max . 50 % at
40 °C
No
heat from heating systems or direct sunlight
No
electromagnetic fields
* For conformity-assessed (verified) balances in accordance with EU requirements, refer to the information on the balance .
** For conformity-assessed (verified) balances in accordance with EU requirements, the legal regulations apply .
15.2 Contamination Type, Overvoltage Category (Device)
Unit
Value
Pollution level according to IEC 61010-1
2
Overvoltage category according to IEC 60664-1
I
15.3 Power Supply
15.3.1 Power Supply Device
Unit
Value
Input voltage
V
DC
15 (
±
10 %)
Power consumption, max .
W
4
Only by Sartorius power supply unit YEPS01-15V0W