− 22 −
Accuracy calculation method
Example) ACV (AC voltage) measurement
Displayed value : 500 V
Range and accuracy : ±(2.0 %rdg + 3 dgt) in 1000 V range
Error : ±(500 V x 2.0 % + 3 dgt) = ±13 V
True value : 500 V ±13 V (487 to 513 V)
* In the 1000 V range, the 3 dgt corresponds to 3 V.
The product specifications described in this manual and its
appearance are subject to change without notice for improvement
or other reasons.