III : COMMUNICATION
IM TEMI 770E_C 2nd Edition: Sep. 1. 2003. 14/21
4.2 Functions
FUNCTION is consisted D-Register for operation.
■
PROGRAM running D-Register
D-Reg. Symbols
Descriptions
D0100
SET_PTNO
Set program pattern number
D0121
WAITMD
Set using WAIT function
D0122
TEMP_WZ
Set wait zone of temperature
D0123
HUMI_WZ
Set wait zone of humidity
D0124
WAIT_TM
Set wait time
■
FIX running D-Register
D-Reg. Symbols
Descriptions
D0102
FIX_T_SP
Set temp SP on FIX running
D0103
FIX_H_SP
Set humi SP on FIX running
D0110
OTMSL
Set using OP TIME(0:OFF, 1:ON)
D0111
OTM_H
Set the hour for FIX OP TIME
D0112
OTM_M
Set the min for FIX OP TIME
■
Operation D-Register
D-Reg. Symbols States
Value
Descriptions
RUN 1
PROG/FIX
RUN
HOLD 2
HOLD
ON/OFF
STEP 3 SEGMENT
STEP
D0102 MODE
STOP 4
PROG/FIX
STOP
PROG 0
PROG
MODE
SETTING
D0104 OPMODE
FIX 1
FIX
MODE
SETTING
STOP 0 STOP
MODE
SETTING
COLD 1
COLD
MODE
SETTING
D0105 PWRMODE
HOT 2 HOT
MODE
SETTING
For PROG RUN's (or FIX RUN) through communication,
TEMI 770 must be STOP(PROG STOP/FIX STOP).
Ex) TEMI 770 should be PROG STOP(D0104 = 0001, D0102 = 0004)
for switching FIX RUN TO PROG RUN.