![Samwontech TEMI770 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/samwontech/temi770/temi770_manual_3793449026.webp)
II.USERS
IM TEMI 770E_C 2nd Edition: Sep. 1. 2003. 7/40
5. MAIN
5-1. FUNCTION
Select
FUNC
button at the MAIN page.(Figure3-4)
Figure 5-1-1. FUNCTION Figure 5-1-2. FIX OP TIME
①
OPERATION : Select PROGRAM Mode or FIX mode.
▶
PROG : Switch to
4.1 PROGRAM STOP
page.
▶
FIX : Switch to
4.3 FIX STOP
page.
②
POWER MODE : Setup when re-start after power failure.
Recognize to power failure if it is more than 3 seconds.
If re-start before 3 seconds, it will run to Hot mode automatically.
▶
STOP : After power failure, go to default.
▶
COLD : After power failure, go to RUN on fix running or go to SEG1 on program running.
▶
HOT : After power failure, running by previous data before power failure.
③④
TEMP(HUMI) SLOPE : Temp/Humi. value goes to the SP gradually.
It is applied both of up and down.
⑤
AT KEY : Select HIDE or DISPLAY for T-AT and H-AT Key on FIX RUN page.
▶
HIDE : Invisible T-AT and H-AT Key.
▶
DISPLAY : Visible T-AT and H-AT Key.
⑥
KEY LOCK : Key lock / unlock.
▶
ON : Key lock (Disable to input data)
▶
OFF : Key unlock (Able to input data)
⑦⑧
FIX OP TIME : Set hour/Set minute.
⑨
FIX OP TIME : Set FIX OP TIME ON/OFF.
FIX OP TIME is for reserving FIX running within 9999 hours 59 min.
It is stop after running time automatically.
⑩
FUSSY SEL : Control over shoot by fussy sel.
⑪
BLGT TIME : Set time of back light.
Figure 5-1-2 FIX OP TIME