44
4.13.2
Cursor Measure
Click “Cursor” in main menu.
This method allows you to take measurements by moving the cursors.
1)
Source
The user can set the source to CH1, CH2 and MATH.
When you use cursors, be sure to set the Source to the waveform on the display that you want to measure.
2)
Type
There are four types of cursors: Cross, Trace, Vertical and Horizontal.
3)
Cross
The Cross cursors appear as cross lines on the display and measure the vertical and horizontal parameters.
Do these steps:
1. Click “Cursor”-> “Source”, select CH1 (select CH2 if you want measure CH2).
2. Click “Cursor”-> “Type”, select Vertical.
3. Push left mouse button, and the vertical lines appear.
4. Drag the mouse button to the point you want to measure.
5. Release the left mouse button, the frequency difference and time difference will be shown at the status bar.
The Cross cursor display window:
The Cross measure result displays on status bar:
4)
Trace
The Trace cursors appear as vertical lines on the display and measure the waveform amplitude at the point the
waveform crosses the cursor.