11.3
Place the eCLS into the case
1. Ensure that a fully charged battery is inserted in the eCLS (see Section 11.4; Figure 11.5).
2. Open the eCLS case with the lead exits pointing up. Place the eCLS into the orange seal
side of the case, with the eCLS battery facing up and the leads pointing up over the lead
exits (see Figure 11.5).
3. Place the leads into the slots in the seal at the lead exits.
Figure 11.5: Place the eCLS with leads connected into the eCLS Case.
4. Close the eCLS case by folding the case shut until it snaps closed (Figure 11.6). Ensure
that the leads do not move from the slots in the seal. The leads may be damaged when
the case is closed if they are not in the slots in the seal at the lead exits.
Figure 11.6 eCLS with two leads connected enclosed in case.
5. After the eCLS is connected and in the eCLS case it may be programmed (see Section 6)
and set up for the patient to wear for the trial stimulation period (see Section 11.5).
11.4
Charge the eCLS battery
1. Before the patient goes home with an eCLS, a fully charged eCLS battery should be in
Orange seal
Lead exits
Battery