![SALUDA MEDICAL Evoke SCS System Скачать руководство пользователя страница 35](http://html2.mh-extra.com/html/saluda-medical/evoke-scs-system/evoke-scs-system_clinical-manual_3627820035.webp)
Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 35 of 101
to that program slot appear in the box.
2. Select electrodes and change settings as required.
Note:
If a program on the CPA has not been run on the stimulator (such as a new
program, or one that has been modified), it will be shown in red in the corresponding
program slot. Once a program has been run, its number will be shown in green (see
Figure 6.24), and it will be saved to the stimulator.
Program 3 not saved
All programs saved
Figure 6.24: (Left) Program 1 and 2 are saved on the stimulator while program 3 is red
because it is not saved to the stimulator. (Right) Program 3 has been run and is now green; all
three programs are currently saved on the stimulator.
6.15.2
To copy settings from one program to another
1. Drag the program number to the desired program slot for the new program.
Note:
You should save the programs to a save slot (see Section 6.15.3) before
overwriting a program in case you want to restore a previous one.
6.15.3
To save a program into the save slots
After the CPA connects or disconnects from a stimulator, all the programs in the program slots
are saved to a save slot and also saved as a program file on the CI.
You can manually create a backup of programs from the stimulator or save a newly created
program for later use in the save slots. This also stores a copy of the programs as a file on the
CI, which can be retrieved in a future programming session (see Section 6.15.4).
1. Touch an empty
save slot
and select
Save
to save all the current programs to that slot
(see Figure 6.23).
Note
: If a save slot that has already been used is selected, the existing program(s) saved
in that slot will be overwritten and a new program file will be saved on the CI, however,
this does not overwrite the previous program file.