Figure 6.8: Settings menu with Patient Zone dropdown menu highlighted.
6.5
Measure electrode impedances
Measure the electrode impedances to determine if all connections are intact between leads,
any lead adapters, and the stimulator.
1.
Touch the
Impedance
icon
.
Note:
The icon will spin during the measurement and stop once it is complete.
2.
The impedances on the 24 electrodes and CLS case (E25) will be shown in the electrode
display window (see Figure 6.9).
Patient Zone
dropdown menu