Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 15 of 101
Figure 6.3: Location of the Visit Name dropdown menu and Connect to Implant button.
7.
Touch the stimulator serial number that you wish to connect to.
8.
The CPA will connect to the stimulator and retrieve any patient data and programs.
Note:
Depending on the amount of data saved on the device this could take several
minutes. Downloading can be cancelled but this is not recommended as any system
error and device usage logs will be missed.
9.
The CPA will display the electrode display window when it has successfully connected
and downloaded data from the stimulator.
Note:
When the CPA connects to a stimulator it will attempt to download three log files
from the stimulator. If any logs cannot be downloaded an error message will be
displayed on the ECAP amplitude and current rolling plot (see Section 8.7). To download
the logs manually, such as when download fails, refer to Section 8.4.
6.2
Set stimulator time
The stimulator time is automatically synchronized to the CI time during connection.
1.
Check that the time and date on the CI are correct (top left corner of CPA screen, Figure
6.2). Refer to the CI instructions (see Section 5.1 for link) to change the time and date.
2.
Touch the
CLS
icon
to open the CLS menu.
3.
Touch
Sync Time
in the CLS menu to set the stimulator time to the CI time (if the time is
Visit Name
dropdown menu
Connect to
Implant button