Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 75 of 101
Figure 11.3: Placing the top cover over the lead and sliding onto lead adapter until it clips into
place.
11.2
Connect the lead adapter to the eCLS
Figure 11.4: Connecting the lead adapter to the eCLS. Port 2 (electrodes 13–24) is shown here.
Port 1 (electrodes 1-12) is on the opposite side of the eCLS.
1. Line up the end of the lead adapter plug with the socket on the eCLS port to ensure the
connector pins do not get bent (Figure 11.4). Push the lead adapter plug into the eCLS
until it clips into place.
a. The eCLS ports are labelled ‘1’ on one side for electrodes 1-12 and ‘2’ on the
other side for electrodes 13-24.
Note:
If you are reconnecting leads to an eCLS that has already been programmed,
ensure that the lead and lead adapter are reinserted into the same ports as used during
previous programming sessions.
2. To connect a second lead to the lead adapter, repeat the steps in Sections 11 and 11.2.