Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 23 of 101
A: Up arrow
B: Maximum current
C: Current scale
D: Current increment scale
E: Desired current level
F: Current increment
G: Down arrow
Figure 6.13: Features of the stimulation current control.
6.9.1
Stop
Touch the
Stop
button
to stop stimulation at any time.
6.10
Set up ECAP measurement
The stimulator records the ECAP signal and displays a plot of the ECAP on the programming
screen (Figure 6.14B). ECAP measurement is only possible for stimulation frequencies of 200 Hz
or less (250 Hz or less for Amplitude option, see Section 6.10.1). ECAPs can only be generated
and measured above a minimum, threshold current, which varies with different postures and
between patients.
A
B
E
G
F
D
C