![SALUDA MEDICAL Evoke SCS System Скачать руководство пользователя страница 47](http://html2.mh-extra.com/html/saluda-medical/evoke-scs-system/evoke-scs-system_clinical-manual_3627820047.webp)
Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 47 of 101
5. Repeat Step 4 for each remaining stimulation set in the program until all have been set
to the threshold level.
6. Set the stimulation sets to run in
Grouped
mode (
),and increase (all stimulation
sets) to the patient preferred level (Figure 7.8).
7. Measurement is only enabled for the applied stimulation set (highlighted in orange. See
Figure 7.8). The ECAP measurement and closed-loop settings from the applied
stimulation set are used for closed-loop control of all stimulation sets.
a. To change the applied stimulation set, select the required stimulation set, then
choose
Apply FB to this SS
in CLS menu – More (Section 8.4).
8. Enable closed-loop stimulation as per Section 6.13.
Note:
Closed-loop stimulation will only operate based on the settings in the applied
stimulation set. The closed-loop control will vary the current to each stimulation set
such that the ratio of current between stimulation sets remains constant.
Figure 7.8: A program with 4 Stimulation Sets configured at desired levels. Note that the
electrode settings from all Stimulation Sets are visible in the background, while the active set
is displayed as normal.
7.5.2
Enable/Disable a stimulation set in a program
To enable/disable a stimulation set in a program:
1. Select the stimulation set you want to enable/disable.
Note:
The
applied stimulation set cannot be disabled.
2. Press
to enable the stimulation set or
to disable it. A disabled stimulation
set will have a red diagonal line across it, e.g.
.
3. Start stimulation again when the desired changes have been made.