Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 66 of 101
Navigation Menu
Open the navigation menu.
Stop
Stop stimulation
Engineering
Information
Used by Saluda Medical staff to obtain
device related information
Factory Reset
Restore devices to default factory
settings. For use by Saluda Medical Field
Clinical Engineer (FCE).
Reset of Last Resort
For use by FCE.
Enter Stock Mode
Stop all CLS functions to conserve
battery. Use the Charger to leave stock
mode.
Visit Name
Set label name for programming visit
Patient Zone
Set the region or zone for the patient
Stim Set Parameters
Set Defaults
Set settings menu values to defaults for
the selected stimulation set
Stimulation Set
selection
Stimulation set slots, labeled A, B, C or
D. Empty slots represent unused
stimulation sets
Pulse Shape
Biphasic - Positive first, Negative first
Triphasic - Positive first, Negative first
Interphase Gap
20 – 500 μs
Program Parameters
Min Inter-Stimulus
Interval
Interval of stimulation set pulses
Auto (minimum permitted ISI),
1400 – 50000 μs
Max Stimulation
Voltage
7.5 – 15 V (ClearCAP disabled)
15 V only (ClearCAP enabled)
Max Stimulation
Ramp Rate
Only applies when connected to CPA.
0.05 – 5.0 mA/s
ECAP Amplifier
Setting
ECAP amplifier gain setting.
High (± 1.8 mV), Low (± 7.2 mV)
Sampling Delay
100 – 1000 μs