5. Troubleshooting
81
Error Messages and Types of Errors
Error
code
Overview
Displayed
message
Description
Instrument operation
when the error occurs
Solution
E001
Invalid
nonvolatile
memory
Unable to
restore setting
data.
The set data in the non-
volatile memory was not
valid.
O p e r ation c ontinue s
based on the factory-set
data. If the factory-set
data is invalid, operation
c o n t i n u e s u s i n g t h e
default value.
C o n t a c t t h e S a k u r a
instrument distributor or
representative.
E002
The operation
status has been
initialized
T he op e r ating s t atu s
data in the non- volatile
memory was not valid.
O p e r ation c ontinue s
i n t h e m a n u a l m o d e
(“Paused” state).
I f t h e p r o b l e m p e r s i s t s
a f t e r r e c o n n e c t i n g t h e
power, contact the Sakura
instrument distributor or
representative.
E003
The record has
been initialized
The log data in the non-
volatile memory was not
valid.
Operation continues.
I f t h e p r o b l e m p e r s i s t s
a f t e r r e c o n n e c t i n g t h e
power, contact the Sakura
instrument distributor or
representative.
E004
Power outage
Power outage
occurred
The power was cut off
while the instrument was
preparing or operating.
(If the power is cut off
or the plug is pulled out
wi tho u t s to p p ing th e
operation, the instrument
r e c o g n ize s a s p owe r
outage.)
O p e r ation c ontinue s
f r o m t h e O p e r a t i o n
screen. If the light was
o n w h e n t h e p o w e r
was cut off, operation
continues with the light
turned off.
Check the molten state of
paraffin, and then resume
using the instrument.
E010
Caution
for high
temperature
Temperature is
too high.
T h e t e m p e r a t u r e o f
each location is too high
c o m p a r e d to t h e s e t
temperature.
If the problem location
is the paraffin chamber,
paraffin line, dispenser
or dispensing solenoid
valve, paraffin cannot
b e disp e nse d. If the
p r o b l e m l o c a t i o n i s
the left/right warming
chamber or left/right hot
plate, normal operation
continues.
Wait until the temperature
of the applicable location
displayed on the Operation
screen stops blinking and
changes to a steady light.
Wait for a while, and if the
temperature does not stop
blinking, contact the Sakura
instrument distributor or
representative. (Check if any
equipment that generates
s t r o n g e l e c t r o m a g n e t i c
waves is nearby.)
High temperature molten
paraffin was added.
E011
Low
temperature
caution
Temperature
has fallen
T h e t e m p e r a t u r e o f
each location is too low
c o m p a r e d to t h e s e t
temperature.
If the problem location
is the paraffin chamber,
paraffin line, dispenser
or dispensing solenoid
valve, paraffin cannot
b e disp e nse d. If the
p r o b l e m l o c a t i o n i s
the left/right warming
chamber or left/right hot
plate, normal operation
continues.
Use the instrument at an
ambient temperature within
10ºC to 35ºC.
Close the paraffin chamber
lid it is open.
Close the warming chamber
lid if it is open.
R e m ove th e t r ay, e tc. i f
placed on the hot plate.
Wait until the temperature
of the applicable location
displayed on the Operation
screen stops blinking and
changes to a steady light.
Wait for a while, and if the
temperature does not stop
blinking, contact the Sakura
instrument distributor or
representative.
(Check if any equipment
t h a t g e n e r a t e s s t r o n g
electromagnetic waves is
nearby.)
S o l i d p a r a f f i n w a s
added.
Unmelted paraffin has
sunk down in the paraffin
chamber.
A cassette basket was
placed in the warming
chamber.
Something (e.g. small
tray or paraffin drain tray
which contains unmelted
paraffin) is placed on the
hot plate.
Ambient temperature is
too low.
The paraffin chamber lid
is open.
The warming chamber
lid is open.
Содержание Tissue-Tek TEC 6
Страница 5: ...Table of contents 5 ...
Страница 6: ...1 Introduction ...
Страница 31: ...1 Introduction 31 ...
Страница 32: ...2 Installation ...
Страница 54: ...3 Explanation of Operations ...
Страница 74: ...4 Error Log ...
Страница 76: ...5 Troubleshooting ...
Страница 95: ...5 Troubleshooting 95 ...
Страница 96: ...6 Replacement Parts ...
Страница 98: ...7 Glossary of Terms ...
Страница 102: ...102 ...
Страница 103: ...M01 021E 02 ...
Страница 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...