3.2.3 Changing the Paraffin Flow Rate
Turning the dial 3.5 times counterclockwise from the fully closed position
(achieved by turning the dial all the way in the clockwise direction), opens the
dial fully.
Turn the paraffin flow rate adjustment dial when paraffin is molten (can be
dispensed).
If the dial is forcibly turned when the paraffin is solid, failure may result.
The paraffin flow rate for dispensing can be adjusted using the paraffin flow
rate adjustment dial.
Continue turning the dial, and it will come off. The dial need not be removed
in normal conditions, but if it was removed, be careful not to lose the spring
inside the dial.
CAUTION:
Do not close the paraffin flow rate adjustment dial of the
Embedding Module fully after the process. If the paraffin
cools down and solidifies when the flow rate adjustment dial
is fully closed, the paraffin may expand the next time it is
melted, and leak out from the flow rate adjustment dial.
パラフィン流量調整ダイヤル
Paraffin flow rate adjustment dial
パラフィン流量調整ダイヤル
スプリング(バネ)
ワッシャー
Paraffin flow rate adjustment dial
Washers
Spring
CAUTION
Содержание Tissue-Tek TEC 6
Страница 5: ...Table of contents 5 ...
Страница 6: ...1 Introduction ...
Страница 31: ...1 Introduction 31 ...
Страница 32: ...2 Installation ...
Страница 54: ...3 Explanation of Operations ...
Страница 74: ...4 Error Log ...
Страница 76: ...5 Troubleshooting ...
Страница 95: ...5 Troubleshooting 95 ...
Страница 96: ...6 Replacement Parts ...
Страница 98: ...7 Glossary of Terms ...
Страница 102: ...102 ...
Страница 103: ...M01 021E 02 ...
Страница 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...