2.2.5 Setting the Auto Panel Display Off Timer
Set the desired time to elapse from the last screen touch before automatically
turning off the backlight on the control panel display.
The panel display turns off only while the operation is paused after elapsing
the scheduled end time in the Auto mode or when not operating by "Heat" or
"Cryo" button in the Manual mode.
1. Touch the “Main Menu” button [1] on
the Operation screen.
2. Touch the “Maintenance” button [2]
on the Main Menu screen.
3. Touch the “Auto Panel Display Off”
button [3] on the Maintenance Menu
screen.
4. If you do not want the panel display
to turn off automatically, select “Not
In Use.”
If you want the panel display to turn
off automatically, you can preset the
panel display switch off timer.
The timer can be set to one of the
following:
1 minute
10 minutes
30 minutes
1 hour
2 hours
When a timer setting is selected, “
•
“ changes to “
•
”.
After the selection, touching the "
" button saves the setting, after which
the display returns to the Maintenance Menu screen.
Touch the " " button, and the display will return to the Maintenance
Menu screen without the setting being saved.
[1]
[2]
[3]
Содержание Tissue-Tek TEC 6
Страница 5: ...Table of contents 5 ...
Страница 6: ...1 Introduction ...
Страница 31: ...1 Introduction 31 ...
Страница 32: ...2 Installation ...
Страница 54: ...3 Explanation of Operations ...
Страница 74: ...4 Error Log ...
Страница 76: ...5 Troubleshooting ...
Страница 95: ...5 Troubleshooting 95 ...
Страница 96: ...6 Replacement Parts ...
Страница 98: ...7 Glossary of Terms ...
Страница 102: ...102 ...
Страница 103: ...M01 021E 02 ...
Страница 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...