3. Explanation of Operations
57
6. Add approx. 80 g into the warming chamber, and then insert the large tray.
This paraffin is used to keep the distribution of tray temperatures uniform.
Add enough paraffin to uniformly cover the bottom of the tray.
NOTES:
• In the figure above, the right warming chamber is set up. The left
warming chamber can be set up in the same manner.
• When the optional small tray is used, the two small trays need to
be placed in one warming chamber. Placing only the one small
tray may cause thermal conduction to slow.
7. It is posiible to use paraffin in the warming chamber tary to soak cassettes
containing tissues.
Add paraffin to the graduation line marked on the warming chamber tray.
The tray containing paraffin is called the “Tissue Tray.”
目盛り線
まで入れる
検体トレイ
CAUTION:
Do not overfill the tray. Doing so may cause paraffin to spill
out, resulting in instrument failure and/or accident.
Add paraffin
to the
marked line.
Tissue tray
CAUTION
Содержание Tissue-Tek TEC 6
Страница 5: ...Table of contents 5 ...
Страница 6: ...1 Introduction ...
Страница 31: ...1 Introduction 31 ...
Страница 32: ...2 Installation ...
Страница 54: ...3 Explanation of Operations ...
Страница 74: ...4 Error Log ...
Страница 76: ...5 Troubleshooting ...
Страница 95: ...5 Troubleshooting 95 ...
Страница 96: ...6 Replacement Parts ...
Страница 98: ...7 Glossary of Terms ...
Страница 102: ...102 ...
Страница 103: ...M01 021E 02 ...
Страница 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...