22
After the 0,5mm scale the beeps shall be short and frequent reaching the ideal level “APEX”, where
this word will be blinking on the panel, when the tip of the instrument reaches the foramen.
ATTENTION: If the instrument outreaches the apex, the word “OVER” (which represents the exceeding
area of the apex) will be blinking followed by intermittent beeps.
The labial clip can be disconnected from the instrument and connected again during the measurement
cycle. It can be exchanged for one or another more adequate during the root canal treatment or
when another canal is measured. In these cases the equipment automatically detects that a new
measurement has been initiated.
NOTE
- The length of work on the canal varies from case to case due to
different forms of the teeth and canals of the dental roots;
- After the length of work has been defined turn off the equipment
and disconnect all cables.
With this function enabled, when pressing the command pedal (a) or button (b) on the own
handpiece the motor will spin clockwise. If, during its operation, the force reaches the predefined
limit concerning the torque, two beeps will sound and the handpiece will automatically spin in
the reverse direction. Once the force is eliminated, the handpiece starts to spin clockwise again.
Activating the REVERSE mode, beeps will sound alerting the alteration in the spinning direction,
spinning it counterclockwise, when disactivating and activating again the equipment the spinning
direction will go back to the initial clockwise direction (NORMAL).
a
b
M 2
AUTO REV
Function (M2) – Inversion of the Automatic Spinning
2.7
N.cm
EQUIPMENT OPERATION
Load within the torque
limit value
More load beyond the
configured torque limit value
If the applied load is
maintained, the file will spin
inversely (reverse).
When the load is removed
Forward spinning
Reverse spinning
Forward spinning