21
Fixing the file on the support
- Press the thumb to the direction of the arrow (A) and fasten the file (B) and release it (C).
a
b
c
On the LCD panel the levels of the scale
of the canal length are divided in the
colors yellow, blue and red.
As soon as the area becomes visible on
the canal apex the yellow color will blink,
alerting the start of the measurement.
The scale is divided in 3 levels, where
each level represents 0,1 mm of the apex
area. The spare yellows represent the
narrow area of the canal apex where
they are defined in 5 levels and can be
adjusted.
When you reach the apex, slowly spin the file counterclockwise until the
meter reaches again 0.5 mm. When the scale of length of the narrow area
on the LCD Display is in 0,5 mm the file will have reached the desired
position, i.e., in the area of the root apex.
Position the rubber stop on the tooth surface as a reference point to
determine the length of work of the root canal. Reminding that as soon
as the file reaches the 2,0 mm position of the canal apex, continuous
beeps will sound.
M 1
M 1
EQUIPMENT OPERATION
early measurement (blue scale)
Press the SET key, the apex mode will blink; the last position of the apex configured will blink, adjust
the apex position in a 0,0 to 1,0mm range by pressing the keys “” and “ADJUST-”. Three (3)
seconds after the last touch the apex mode will stop blinking and the configuration will be saved.
Generally the narrow area is between 0.2 and 0.3mm of the canal apex. By initiating the procedure
the equipment will detect the beginning of the measurement.
By initiating the procedure the equipment will show the distance between the tip of the file (A)
and the canal apex (B), according to scale below on the LCD panel.
FLASHING
(yellow scale)
(red scale)
M1
Support
file
Stop
rubber