16

2. Lubrication

■ Lubrication product list

• Angle handpiece

① Insert the spray nozzle to motor assembly part (the end part of angle) and spray lubricant 2~3 seconds.

② Insert the nozzle tip to head part and spray lubricant 2~3 seconds. 

   - Do not lubricate the motor section of the motor handpiece. It may cause bearing failure and

     heat generation. 

• The nozzle tip and spray nozzle are not provided by manufacturer.

• If the continuous operation, please put the oil every an hour.

• When lubrication, please hold the handpiece firmly because the handpiece can go out by the spray pressure.

Spray Nozzle

Nozzle

tip

3. Sterilization

■ Sterilization product list

• Angle handpiece, Lip hook, file clip

■ Sterilization Method

① Dry the applicable product completely after cleaning.

② Put the applicable product in a sterile bag and seal tightly.

③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃, and let it dry for 30 minutes. 

Temperature

132℃

Time

4 minute

Drying time

30 minute

Sterilization

Moist heat sterilization

- Maximum sterilization number is 250 times.

- Moist steam sterilization only. Otherwise the products might be damaged.

- Make sure to use the sterilization bag and seal tightly.

- Make sure to dry the applicable products before sterilization.

- Sterilizing the angle handpiece with blood or other remains may cause the breakage of the

   angle handpiece.

   Therefore, it is necessary to clean the angle handpiece before sterilization.

- Keep dry a connector and socket of the control box.

- Before using, dry the product completely.

132

4. Battery Replacement

E-CUBE uses rechargeable battery. Battery is reused about 300 ~ 500 cycles, the span life may be different according

to handpieces. Replace the battery when it shows shorter operation and recharging time and weak torque than set up.

Be aware of ‘Caution’ for battery replacement. Manufacturer is not responsible for damage or breakage of the product

caused by not following the instructions ‘Caution’ for changing battery.

Spec : 12Vd.c, 800mAh

Содержание E-CUBE PLUS

Страница 1: ...INSTRUCTION MANUAL www saeshin com...

Страница 2: ...ES PRODUCT COMPONENTS PART NAME OF CONTROL BOX BUTTONS ON OPERATION PANEL LCD WINDOW CONNECTION OF EACH PART OPERATION PROCESS MAINTENANCE OF THE PRODUCT ATTENTION PRODUCT SPECIFICATION ELECTROMAGNETI...

Страница 3: ...ainst electric shock Class equipment To avoid the risk of electric shock the product must be connected to a power source with a protective ground Degree of protection against electric shock Type B app...

Страница 4: ...ident or failure User has responsibility to operate and maintain the product This product uses a rechargeable battery Ni MH battery Recycling is possible but some countries may not allow processing so...

Страница 5: ...aker If patient is taking antibiotics patient needs to consult with doctor before operation Keep away from fire or heat when using spray containing flammable material Manufacturer does not have any re...

Страница 6: ...apex measurement starts the motor starts to rotate automatically and when the measurement is finished the motor stops Available only in APEX ENDO mode AUTO SLOW DOWN The rotational speed of the motor...

Страница 7: ...S No No Control Box Adaptor and AC Power Cord Angle Handpiece Option Motor Option Motor Stand Foot Controller Battery Including in the control box COMPONENTS Probe Cord Lip Hook File Clip Connecting C...

Страница 8: ...or reverse rotation when changing the rotation angle Gear Ratio Button Use for setting the gear ratio Speed Rotation Angle Button Use for setting the speed RPM Press and hold the button can set the sp...

Страница 9: ...O mode it shows the load condition of motor during operating 6 Speed Display It shows the preset RPM In reciprocating mode programs 8 and 9 the currently set angle is displayed when selecting the forw...

Страница 10: ...8 Adapter connector Foot controller connector Battery cover Probe connector Motor connector CONNECTION OF EACH PART...

Страница 11: ...product at consistent temperature location Please insert the pins of motor cord to the holes position of control box carefully 1 Motor Connection 2 Probe Cord Connection Connect the file clip and lip...

Страница 12: ...the connecting cord into the file clip holder with 3 o clock or 9 o clock directions and rotate 90 to fix it Connect the connecting cord to the probe cord then insert it into the cord clip Secure the...

Страница 13: ...te the angle handpiece and motor pull out the angle handpiece to its axis side 6 File Connection Please insert the file into the chuck of angle handpiece until stop the file To separate the file and a...

Страница 14: ...ously applied to the motor it may automatically stop to prevent overheating In this case leave the motor for a while until it cools down When the display returns to normal the product can be used Auto...

Страница 15: ...ot controller Motor Button is of motor the motor stop operation by pressing the motor button once more Motor operation can be controlled by pressing and releasing the foot controller 7 After use place...

Страница 16: ...ck the length of the measured root canal and the length by taking an x ray OPERATION PROCESS APEX MODE 1 Turn on the control box by pressing the power button over the 3 seconds 2 Press APEX ENDO butto...

Страница 17: ...le only in APEX ENDO mode MAINTENANCE OF THE PRODUCT 1 Cleaning Angle Handpiece Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains blood physiologic saline soluti...

Страница 18: ...and let it dry for 30 minutes Temperature 132 Time 4 minute Drying time 30 minute Sterilization Moist heat sterilization Maximum sterilization number is 250 times Moist steam sterilization only Otherw...

Страница 19: ...right after charging there may be an abnormality Contact the manufacturer The adaptor fuse is blown out Check the connection The adaptor is not connected to the control box Check the connection The a...

Страница 20: ...duct 1 Do not put any physical pressure on the adaptor cord motor cord and foot controller cord 6 Product Disposal Follow your country specific law directives standards and guidelines for the disposal...

Страница 21: ...ame AC 100 240V 47 63Hz 0 9 0 34A DC 20 V 1 5A Max Sinpro Electronics Co Ltd MPU31 107 1 Control Box 2 Motor 3 Angle 5 Adaptor Name Gear ratio Bur replacement type Bur dimension Chuck type ACL B 51EPB...

Страница 22: ...rom that to which the other device s are connected Consult with the dealer Electromagnetic Emissions The E CUBE Plus is intended for use in the electromagnetic environment specified below The customer...

Страница 23: ...cle at 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for...

Страница 24: ...o AM FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurem...

Страница 25: ...objects and persons 150 kHz 80 MHz d 1 2 P Separation distance according to frequency of transmitter m 800 MHz 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 1...

Страница 26: ...use or improper use and treatment Use the product with incorrect input voltage AC voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction aft...

Страница 27: ...clave Type B This product should not be mixed with other commercial waste for disposal Non sterile Fuse Speed Reciprocating Angle Ratio Button Torque Button Program Button Apical Line Button Volume Bu...

Страница 28: ...53 580 0999 1 52 5 Tel 053 587 2345 Fax 053 580 0918 SOM S007S E 2022 01 27 Rev 0 www saeshin com traus saeshin com MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park...

Отзывы: