2

USER GUIDELINES

CAUTION

• Please pay attention to the use considering the patient’s safety first of all.

• Check condition of the product before use.

• Stop to use and inspect the product when an abnormal sign such as calescence, vibration or noise

     occurs before or during operation.

• Please contact manufacturer if cannot find reason of problem, or cannot solve the problem.

• Replace the bur or file after the motor is completely stopped. Otherwise, inside of the angle

     could be damaged, and it could cause an accident.

• Please sterilize and lubricate the handpiece right after use. Otherwise, the product could be

     damaged due to coagulation of blood inside the product.

• This product is not sterilized when shipped from the manufacturer. 

    Refer to the maintenance section of the product and sterilize components that require sterilization

    before use.

• The control box, motor, foot controller and adaptor are not steam sterilization objects.

• Contact manufacturer if incorrect display is on the screen. There is a risk of an accident.

• Do not drop or give a strong impact to the product. It may cause a breakdown.

• Do not insert oil into the motor. It will damage bearing or cause calescence.

• The product is designed to be used indoor, do not use outdoor.

• Do not use the product in humid or dusty environment.

• Use the regulated products and consumables by manufacturer. Otherwise there might be a risk of

     an accident or failure.

• User has responsibility to operate and maintain the product.

• This product uses a rechargeable battery [Ni-MH battery]. Recycling is possible, but some countries

     may not allow processing, so return it to the place of purchase.

• When connecting the cord and connector, check the direction and attach it.

• When the battery bar is one level on the screen, please charge it as much as possible.

• If the battery has been almost discharged, the battery level on the screen is blank and a beep sounds.

     Stop using it immediately and charge it..

• When using apex measurements, check if the patient is allergic to metals.

• Excessive force on the root canal may result in damage or blockage of the root canal wall caused by the file.

• Be careful when using reverse rotation, as the torque of the motor is strong and the file may be broken.

WARNING

• Please watch old person, children, or disabled person near the place that the product is installed.
     If so, do not let them alone.
• Portable and mobile radio frequency communications equipment may affect the product.
     Do not use radio frequency equipment in close proximity for the product.
• Before measuring root canal length, make sure the "Auto Start/Stop" function is turned off.
     If it is on, the motor will rotate automatically.
• If the root canal is blocked or under special circumstances, the measurement may not be
     possible or accurate.
• Do not let the metal part of the contra-angle come into contact with the oral mucosa.
     There is a risk of an accident due to the rotation of the motor.
• Do not modify this equipment without authorization of the manufacturer.

Содержание E-CUBE PLUS

Страница 1: ...INSTRUCTION MANUAL www saeshin com...

Страница 2: ...ES PRODUCT COMPONENTS PART NAME OF CONTROL BOX BUTTONS ON OPERATION PANEL LCD WINDOW CONNECTION OF EACH PART OPERATION PROCESS MAINTENANCE OF THE PRODUCT ATTENTION PRODUCT SPECIFICATION ELECTROMAGNETI...

Страница 3: ...ainst electric shock Class equipment To avoid the risk of electric shock the product must be connected to a power source with a protective ground Degree of protection against electric shock Type B app...

Страница 4: ...ident or failure User has responsibility to operate and maintain the product This product uses a rechargeable battery Ni MH battery Recycling is possible but some countries may not allow processing so...

Страница 5: ...aker If patient is taking antibiotics patient needs to consult with doctor before operation Keep away from fire or heat when using spray containing flammable material Manufacturer does not have any re...

Страница 6: ...apex measurement starts the motor starts to rotate automatically and when the measurement is finished the motor stops Available only in APEX ENDO mode AUTO SLOW DOWN The rotational speed of the motor...

Страница 7: ...S No No Control Box Adaptor and AC Power Cord Angle Handpiece Option Motor Option Motor Stand Foot Controller Battery Including in the control box COMPONENTS Probe Cord Lip Hook File Clip Connecting C...

Страница 8: ...or reverse rotation when changing the rotation angle Gear Ratio Button Use for setting the gear ratio Speed Rotation Angle Button Use for setting the speed RPM Press and hold the button can set the sp...

Страница 9: ...O mode it shows the load condition of motor during operating 6 Speed Display It shows the preset RPM In reciprocating mode programs 8 and 9 the currently set angle is displayed when selecting the forw...

Страница 10: ...8 Adapter connector Foot controller connector Battery cover Probe connector Motor connector CONNECTION OF EACH PART...

Страница 11: ...product at consistent temperature location Please insert the pins of motor cord to the holes position of control box carefully 1 Motor Connection 2 Probe Cord Connection Connect the file clip and lip...

Страница 12: ...the connecting cord into the file clip holder with 3 o clock or 9 o clock directions and rotate 90 to fix it Connect the connecting cord to the probe cord then insert it into the cord clip Secure the...

Страница 13: ...te the angle handpiece and motor pull out the angle handpiece to its axis side 6 File Connection Please insert the file into the chuck of angle handpiece until stop the file To separate the file and a...

Страница 14: ...ously applied to the motor it may automatically stop to prevent overheating In this case leave the motor for a while until it cools down When the display returns to normal the product can be used Auto...

Страница 15: ...ot controller Motor Button is of motor the motor stop operation by pressing the motor button once more Motor operation can be controlled by pressing and releasing the foot controller 7 After use place...

Страница 16: ...ck the length of the measured root canal and the length by taking an x ray OPERATION PROCESS APEX MODE 1 Turn on the control box by pressing the power button over the 3 seconds 2 Press APEX ENDO butto...

Страница 17: ...le only in APEX ENDO mode MAINTENANCE OF THE PRODUCT 1 Cleaning Angle Handpiece Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains blood physiologic saline soluti...

Страница 18: ...and let it dry for 30 minutes Temperature 132 Time 4 minute Drying time 30 minute Sterilization Moist heat sterilization Maximum sterilization number is 250 times Moist steam sterilization only Otherw...

Страница 19: ...right after charging there may be an abnormality Contact the manufacturer The adaptor fuse is blown out Check the connection The adaptor is not connected to the control box Check the connection The a...

Страница 20: ...duct 1 Do not put any physical pressure on the adaptor cord motor cord and foot controller cord 6 Product Disposal Follow your country specific law directives standards and guidelines for the disposal...

Страница 21: ...ame AC 100 240V 47 63Hz 0 9 0 34A DC 20 V 1 5A Max Sinpro Electronics Co Ltd MPU31 107 1 Control Box 2 Motor 3 Angle 5 Adaptor Name Gear ratio Bur replacement type Bur dimension Chuck type ACL B 51EPB...

Страница 22: ...rom that to which the other device s are connected Consult with the dealer Electromagnetic Emissions The E CUBE Plus is intended for use in the electromagnetic environment specified below The customer...

Страница 23: ...cle at 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for...

Страница 24: ...o AM FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurem...

Страница 25: ...objects and persons 150 kHz 80 MHz d 1 2 P Separation distance according to frequency of transmitter m 800 MHz 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 1...

Страница 26: ...use or improper use and treatment Use the product with incorrect input voltage AC voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction aft...

Страница 27: ...clave Type B This product should not be mixed with other commercial waste for disposal Non sterile Fuse Speed Reciprocating Angle Ratio Button Torque Button Program Button Apical Line Button Volume Bu...

Страница 28: ...53 580 0999 1 52 5 Tel 053 587 2345 Fax 053 580 0918 SOM S007S E 2022 01 27 Rev 0 www saeshin com traus saeshin com MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park...

Отзывы: