14
1. Turn on the control box by pressing the power button over the 3 seconds.
Press APEX/ENDO button to set APEX mode .
2. Press Apical Line button to change Apical Line value.
3. Hang the lip hook on the patient's lips.
4. The file is inserted into the patient's teeth..
5. Hang the file clip to the file.
6. Checking the graph and measured value on the LCD display, move the file slowly along the root canal,
and the motor automatically stops when the file reach the default apical value(factory default value is
set to 0.5), the apex symbol “APEX”
blinks and the motor automatically stops.
7. After placing the stopper on the file at the tooth reference point, remove the file from the tooth.
8. Cross-check the length of the measured root canal and the length by taking an x-ray.
OPERATION PROCESS (APEX MODE)
1. Turn on the control box by pressing the power button over the 3 seconds.
2. Press APEX/ENDO button to set APEX+ENDO combination mode .
3. Press Apical Line button to change Apical Line value.
(Set the special functions if necessary.)
4. Please insert the file into angle handpiece. (Refer to CONNECTION OF EACH PART)
5. Connect the file clip to the connecting cord of the angle handpiece.
6. Hang the lip hook on the patient's lips.
7. Refer to the OPERATION PROCESS (ENDO MODE), insert the file into the patient's teeth and proceed
with the operation.
8. Checking the graph and measured value on the LCD display, move the file slowly along the root canal, and
the motor automatically stops when the file reach the default apical value (factory default value is set to 0.5).
OPERATION PROCESS (APEX+ENDO MODE)
Содержание E-CUBE PLUS
Страница 1: ...INSTRUCTION MANUAL www saeshin com...