- For more accurate measurement, connect the provided tester to the probe connector of the main
body and check that the apical value is within the range of 0.4 to 0.6.
- AUTO START/STOP & AUTO SLOW DOWN are not supported in the reciprocating mode.
(Program No. 8, 9).
<AUTO START/STOP>
- If you long press the Apical Line button once, is displayed and the AUTO START/STOP function
is executed.
- When the apex measurement starts, the motor starts to rotate automatically, and when the measurement is
inished, the motor stops. (Available only in APEX+ENDO mode)
<AUTO SLOW DOWN>
- If you long press the Apical Line button twice, is displayed and the AUTO SLOW DOWN function
is executed.
- The rotational speed of the motor gradually decreases according to the ratio of the apex measurement depth,
and the motor stops when it reaches the set APICAL LINE. (Available only in APEX+ENDO mode)
MAINTENANCE OF THE PRODUCT
1. Cleaning
■ Angle Handpiece
① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains
(blood, physiologic saline solution, etc.) and rinse.
※ Do not put tap water in the motor. May cause malfunction.
② Disconnect the angle handpiece to the motor.
③ Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes
in Enzymatic Detergent.
④ Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
⑥ Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly
subjected 3 times.
(Change the purified water each time.)
⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.
■ Control box, Motor, Foot controller
Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.
※ Pay attention to not enter the water into the motor.