Note: For the HL-DS-U4 and HL-DS-U4-420, if the analog inputs of the HL-DS are required,
the configuration of the HL-DS can be changed via HW4 and the devices must be
integrated again into RMS or the settings can also be changed via the RMS
(Tools>Setup>Devices>Properties>Show, a window appears where the settings can be
changed. After clicking <Apply> the device need to be deleted from RMS and reintegrated
again to show the new input settings.
IMPORTANT: If not all checkmarks for measurement points are set and it is wished to
expand it later then the same device needs to be added again with the same identification
and IP address (there is no need to delete the device before adding it). The desired check
marks can be set. And afterwards, the added or updated measurement points will appear.
Step 4
The device is added.
Step 5
IMPORTANT: Overwritten settings.
By adding the device to RMS via the RMS-CONVERTER, the following configuration will be
overwritten on the device:
Alarm settings: will not be overwritten, but ignored by RMS. Meaning that local alarms
can be set and the HL-NT will alarm via the display and buzzer.
Log file format: .LOG
Delete old log files: disabled
Relais settings: off
Write protection: off
Current time: UTC
Log interval: 1 min
Create new file: every day
Relative humidity: on
Temperature: on
Calculated value: off
Start time: 1.1.2000
Stop time: 1.1.2050
All .XLS files will be deleted.
All .LOG files older than 7 day will be deleted
Note: if one of the settings above is changed with HW4 further to the initial installation
into RMS, then RMS will not note the change. However, such a change would cause RMS
to not work correctly. The changes will however be overwritten should the RMS-
CONVERTER reboot (due to a short power interrupt).
Содержание RMS-CONVERTER
Страница 1: ...2017 ROTRONIC AG Bassersdorf Switzerland RMS CONVERTER...
Страница 101: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 101 of 126...
Страница 102: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 102 of 126...
Страница 106: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 106 of 126...